=Spectroscopy Setup= = Spectroscopy Setup =

Spectroscopy Setup

Dynamic range per capacitance

  • Procedures : FmIltProc/DynamicRangeSpectro

  • Pre-test setup:
    • Note, that the related PTD contains a combined measurement of 1.2.3, 1.2.8 and 1.2.18 blocks, and cannot be applied in its original form.
    • Set the OGSE BB temperatures (SET_UP_OGSE_BB_TEMP_OBS, see the temperatures below)
    • Move filter wheel to "POS A" (SPEC_fltw_move)
  • CUS scripts: SPEC_dyn_range_obs, SPEC_DiagHK_DynRange_block_stop, SPEC_DiagHK_DynRange_block_start

  • Each measurement block (OGSE BB temperature setup) runs through all combinations of integration capacitor (0.1pF, 0.3pF, 1pF, 3pF) and reset interval values (1/8s, 1/4s, 1/2s, 2s), and is performed for two grating positions (GR-1 = 760000 and GR-2 = 490000).
    • Setup #1: OBSE BB1: 16.3K, OGSE BB2: 15.7K
    • Setup #2: OBSE BB1: 18.8K, OGSE BB2: 18.2K
    • Setup #3: OBSE BB1: 25.4K, OGSE BB2: 24.6K
    • Setup #4: OBSE BB1: 32.5K, OGSE BB2: 31.5K
    • Setup #5: OBSE BB1: 40.7K, OGSE BB2: 39.3K
    • Setup #6: OBSE BB1: 48.0K, OGSE BB2: 46.0K
    • Setup #7: OBSE BB1: 61.4K, OGSE BB2: 58.6K
    • Setup #8: OBSE BB1: 77.0K, OGSE BB2: 73.0K
  • Estimated duration of the tests for each indivitual OGSE BB setting: 2x4x4x128s + 2x15s for grating movement + 1080s OGSE BB stabilization = 1.45h