Size: 5268
Comment:
|
Size: 1345
Comment:
|
Deletions are marked like this. | Additions are marked like this. |
Line 1: | Line 1: |
=Spectroscopy Setup= | = Spectroscopy Setup = |
Line 8: | Line 8: |
* Note, that the related PTD contains a combined measurement of 1.2.3, 1.2.8 and 1.2.18 blocks, and cannot be applied in its original form. | |
Line 11: | Line 12: |
* Below are all combinations of the OGBE BB temperature, integration capacitor and reset interval values. Note, that the parameter of the SPEC_dynamic_range_fmilt123_OBS CUS script is the ramp length, and not the reset interval! (1/16s -> 16; 1/4s -> 64; 1/2s -> 128; 2s -> 512). The integration capacitor/reset interval block is performed for two grating positions: GR-1 = 760000 and GR-2 = 490000. | * Each measurement block (OGSE BB temperature setup) runs through all combinations of integration capacitor (0.1pF, 0.3pF, 1pF, 3pF) and reset interval values (1/8s, 1/4s, 1/2s, 2s), and is performed for two grating positions (GR-1 = 760000 and GR-2 = 490000). |
Line 15: | Line 16: |
|| GR-1 || 0.1pF || 0.3pF || 1pF || 3pF || || 1/16s || || || || || || 1/4s || || || || || || 1/2s || || || || || || 2s || || || || || || GR-2 || 0.1pF || 0.3pF || 1pF || 3pF || || 1/16s || || || || || || 1/4s || || || || || || 1/2s || || || || || || 2s || || || || || |
|
Line 28: | Line 17: |
|| GR-1 || 0.1pF || 0.3pF || 1pF || 3pF || || 1/16s || || || || || || 1/4s || || || || || || 1/2s || || || || || || 2s || || || || || || GR-2 || 0.1pF || 0.3pF || 1pF || 3pF || || 1/16s || || || || || || 1/4s || || || || || || 1/2s || || || || || || 2s || || || || || |
|
Line 44: | Line 20: |
|| GR-1 || 0.1pF || 0.3pF || 1pF || 3pF || || 1/16s || || || || || || 1/4s || || || || || || 1/2s || || || || || || 2s || || || || || || GR-2 || 0.1pF || 0.3pF || 1pF || 3pF || || 1/16s || || || || || || 1/4s || || || || || || 1/2s || || || || || || 2s || || || || || |
|
Line 58: | Line 21: |
|| GR-1 || 0.1pF || 0.3pF || 1pF || 3pF || || 1/16s || || || || || || 1/4s || || || || || || 1/2s || || || || || || 2s || || || || || || GR-2 || 0.1pF || 0.3pF || 1pF || 3pF || || 1/16s || || || || || || 1/4s || || || || || || 1/2s || || || || || || 2s || || || || || |
|
Line 74: | Line 24: |
|| GR-1 || 0.1pF || 0.3pF || 1pF || 3pF || || 1/16s || || || || || || 1/4s || || || || || || 1/2s || || || || || || 2s || || || || || || GR-2 || 0.1pF || 0.3pF || 1pF || 3pF || || 1/16s || || || || || || 1/4s || || || || || || 1/2s || || || || || || 2s || || || || || |
|
Line 88: | Line 25: |
|| GR-1 || 0.1pF || 0.3pF || 1pF || 3pF || || 1/16s || || || || || || 1/4s || || || || || || 1/2s || || || || || || 2s || || || || || || GR-2 || 0.1pF || 0.3pF || 1pF || 3pF || || 1/16s || || || || || || 1/4s || || || || || || 1/2s || || || || || || 2s || || || || || |
|
Line 105: | Line 28: |
|| GR-1 || 0.1pF || 0.3pF || 1pF || 3pF || || 1/16s || || || || || || 1/4s || || || || || || 1/2s || || || || || || 2s || || || || || || GR-2 || 0.1pF || 0.3pF || 1pF || 3pF || || 1/16s || || || || || || 1/4s || || || || || || 1/2s || || || || || || 2s || || || || || |
|
Line 120: | Line 29: |
|| GR-1 || 0.1pF || 0.3pF || 1pF || 3pF || || 1/16s || || || || || || 1/4s || || || || || || 1/2s || || || || || || 2s || || || || || || GR-2 || 0.1pF || 0.3pF || 1pF || 3pF || || 1/16s || || || || || || 1/4s || || || || || || 1/2s || || || || || || 2s || || || || || |
|
Line 135: | Line 31: |
* Measurements / Preliminary analysis : FmIltAnal/DynamicRangeSpectro |
Spectroscopy Setup
Dynamic range per capacitance
Procedures : FmIltProc/DynamicRangeSpectro
- Pre-test setup:
- Note, that the related PTD contains a combined measurement of 1.2.3, 1.2.8 and 1.2.18 blocks, and cannot be applied in its original form.
- Set the OGSE BB temperatures (SET_UP_OGSE_BB_TEMP_OBS, see the temperatures below)
- Move filter wheel to "POS A" (SPEC_fltw_move)
CUS scripts: SPEC_dyn_range_obs, SPEC_DiagHK_DynRange_block_stop, SPEC_DiagHK_DynRange_block_start
- Each measurement block (OGSE BB temperature setup) runs through all combinations of integration capacitor (0.1pF, 0.3pF, 1pF, 3pF) and reset interval values (1/8s, 1/4s, 1/2s, 2s), and is performed for two grating positions (GR-1 = 760000 and GR-2 = 490000).
- Setup #1: OBSE BB1: 16.3K, OGSE BB2: 15.7K
- Setup #2: OBSE BB1: 18.8K, OGSE BB2: 18.2K
- Setup #3: OBSE BB1: 25.4K, OGSE BB2: 24.6K
- Setup #4: OBSE BB1: 32.5K, OGSE BB2: 31.5K
- Setup #5: OBSE BB1: 40.7K, OGSE BB2: 39.3K
- Setup #6: OBSE BB1: 48.0K, OGSE BB2: 46.0K
- Setup #7: OBSE BB1: 61.4K, OGSE BB2: 58.6K
- Setup #8: OBSE BB1: 77.0K, OGSE BB2: 73.0K
- Estimated duration of the tests for each indivitual OGSE BB setting: 2x4x4x128s + 2x15s for grating movement + 1080s OGSE BB stabilization = 1.45h