614
Comment:
|
2374
|
Deletions are marked like this. | Additions are marked like this. |
Line 1: | Line 1: |
=Spectroscopy Setup= | = Spectroscopy Setup = |
Line 3: | Line 3: |
== Bias loops and detector heater loops== | |
Line 5: | Line 4: |
== Dynamic range per capacitance == | == Dynamic range per capacitance (1.2.3) == * Procedures : FmIltProc/DynamicRangeSpectro * Pre-test setup: * Note, that the related PTD contains a combined measurement of 1.2.3, 1.2.8 and 1.2.18 blocks, and cannot be applied in its original form. * Set the OGSE BB temperatures (SET_UP_OGSE_BB_TEMP_OBS, see the temperatures below) * Move filter wheel to "POS A" (SPEC_fltw_move) * CUS scripts: SPEC_dyn_range_obs, SPEC_DiagHK_DynRange_block_stop, SPEC_DiagHK_DynRange_block_start * Each measurement block (OGSE BB temperature setup) runs through all combinations of integration capacitor (0.1pF, 0.3pF, 1pF, 3pF) and reset interval values (1/8s, 1/4s, 1/2s, 2s), and is performed for two grating positions (GR-1 = 760000 and GR-2 = 490000). |
Line 7: | Line 14: |
* Setup #1: OBSE BB1: 16.3K, OGSE BB2: 15.7K * Setup #2: OBSE BB1: 18.8K, OGSE BB2: 18.2K * Setup #3: OBSE BB1: 25.4K, OGSE BB2: 24.6K * Setup #4: OBSE BB1: 32.5K, OGSE BB2: 31.5K * Setup #5: OBSE BB1: 40.7K, OGSE BB2: 39.3K * Setup #6: OBSE BB1: 48.0K, OGSE BB2: 46.0K * Setup #7: OBSE BB1: 61.4K, OGSE BB2: 58.6K * Setup #8: OBSE BB1: 77.0K, OGSE BB2: 73.0K * Estimated duration of the tests for each indivitual OGSE BB setting: 2x4x4x128s + 2x15s for grating movement + 1080s OGSE BB stabilization = 1.45h == Signal dependence on chopper frequency (1.2.8) == |
|
Line 8: | Line 35: |
* Temperature settings for OGSE BB1: 16.3, 18.8, 25.4, 32.5, 40.7, 48.0, 61.4, 77.0 * Temperature settings for OGSE BB2: 15.7, 18.2, 24.6, 31.5, 39.3, 46.0, 58.6, 73.0 * Reset intervals: 1/16s, 1/4s, 1/2s, 2s * Integration capacitors: 0.1pF, 0.3pF, 1.0pF, 3.0pF * CUS scripts: SPEC_dynamic_range_fmilt123_OBS, SPEC_DiagHK_DynRange_block_stop, SPEC_DiagHK_DynRange_block_start * Measurements / Preliminary analysis : FmIltAnal/DynamicRangeSpectro |
* Pre-test setup: * CS1 temperature: 70.5K (fixed) * CS2 temperature: 76.0K (fixed) * Grating: 425250 (fixed) * Reset interval: 1/4s (fixed) * SPU setup: Entire slope fitting * Set the OGSE BB temperatures (SET_UP_OGSE_BB_TEMP_OBS, see the temperatures below) * Move filter wheel to "POS B" (SPEC_fltw_move) * CUS script: SPEC_chopper_freq_fmilt128_obs * Each measurement block (OGSE BB temperature setup) runs performs a tringular chopping measurement between the calibration sources and the OGSE BB1 as OGSE BB1 -- CS1 -- OGSE BB1 -- CS2 -- OBSE BB1 -- CS1 -- ... * Setup #1: OBSE BB1: 18.8K * Setup #2: OBSE BB1: 25.4K * Setup #3: OBSE BB1: 32.5K * Setup #4: OBSE BB1: 40.7K * Setup #5: OBSE BB1: 48.0K * Setup #6: OBSE BB1: 61.4K * Setup #7: OBSE BB1: 77.0K * Estimated duration of the tests for each indivitual OGSE BB setting: ~0.45h |
Spectroscopy Setup
Dynamic range per capacitance (1.2.3)
Procedures : FmIltProc/DynamicRangeSpectro
- Pre-test setup:
- Note, that the related PTD contains a combined measurement of 1.2.3, 1.2.8 and 1.2.18 blocks, and cannot be applied in its original form.
- Set the OGSE BB temperatures (SET_UP_OGSE_BB_TEMP_OBS, see the temperatures below)
- Move filter wheel to "POS A" (SPEC_fltw_move)
CUS scripts: SPEC_dyn_range_obs, SPEC_DiagHK_DynRange_block_stop, SPEC_DiagHK_DynRange_block_start
- Each measurement block (OGSE BB temperature setup) runs through all combinations of integration capacitor (0.1pF, 0.3pF, 1pF, 3pF) and reset interval values (1/8s, 1/4s, 1/2s, 2s), and is performed for two grating positions (GR-1 = 760000 and GR-2 = 490000).
- Setup #1: OBSE BB1: 16.3K, OGSE BB2: 15.7K
- Setup #2: OBSE BB1: 18.8K, OGSE BB2: 18.2K
- Setup #3: OBSE BB1: 25.4K, OGSE BB2: 24.6K
- Setup #4: OBSE BB1: 32.5K, OGSE BB2: 31.5K
- Setup #5: OBSE BB1: 40.7K, OGSE BB2: 39.3K
- Setup #6: OBSE BB1: 48.0K, OGSE BB2: 46.0K
- Setup #7: OBSE BB1: 61.4K, OGSE BB2: 58.6K
- Setup #8: OBSE BB1: 77.0K, OGSE BB2: 73.0K
- Estimated duration of the tests for each indivitual OGSE BB setting: 2x4x4x128s + 2x15s for grating movement + 1080s OGSE BB stabilization = 1.45h
Signal dependence on chopper frequency (1.2.8)
Procedures : FmIltProc/DynamicRangeSpectro
- Pre-test setup:
- CS1 temperature: 70.5K (fixed)
- CS2 temperature: 76.0K (fixed)
- Grating: 425250 (fixed)
- Reset interval: 1/4s (fixed)
- SPU setup: Entire slope fitting
- Set the OGSE BB temperatures (SET_UP_OGSE_BB_TEMP_OBS, see the temperatures below)
- Move filter wheel to "POS B" (SPEC_fltw_move)
- CUS script: SPEC_chopper_freq_fmilt128_obs
- Each measurement block (OGSE BB temperature setup) runs performs a tringular chopping measurement between the calibration sources and the OGSE BB1 as OGSE BB1 -- CS1 -- OGSE BB1 -- CS2 -- OBSE BB1 -- CS1 -- ...
- Setup #1: OBSE BB1: 18.8K
- Setup #2: OBSE BB1: 25.4K
- Setup #3: OBSE BB1: 32.5K
- Setup #4: OBSE BB1: 40.7K
- Setup #5: OBSE BB1: 48.0K
- Setup #6: OBSE BB1: 61.4K
- Setup #7: OBSE BB1: 77.0K
- Estimated duration of the tests for each indivitual OGSE BB setting: ~0.45h