=Spectroscopy Setup= = Spectroscopy Setup =
Line 3: Line 3:
== Bias loops and detector heater loops==
Line 5: Line 4:
== Dynamic range per capacitance == == Dynamic range per capacitance (1.2.3) ==
 
 * Procedures : FmIltProc/DynamicRangeSpectro
 * Pre-test setup:
   * Note, that the related PTD contains a combined measurement of 1.2.3, 1.2.8 and 1.2.18 blocks, and cannot be applied in its original form.
   * Set the OGSE BB temperatures (SET_UP_OGSE_BB_TEMP_OBS, see the temperatures below)
   * Move filter wheel to "POS A" (SPEC_fltw_move)
 * CUS scripts: SPEC_dyn_range_obs, SPEC_DiagHK_DynRange_block_stop, SPEC_DiagHK_DynRange_block_start
 * Each measurement block (OGSE BB temperature setup) runs through all combinations of integration capacitor (0.1pF, 0.3pF, 1pF, 3pF) and reset interval values (1/8s, 1/4s, 1/2s, 2s), and is performed for two grating positions (GR-1 = 760000 and GR-2 = 490000).
Line 7: Line 14:
  * Setup #1: OBSE BB1: 16.3K, OGSE BB2: 15.7K

  * Setup #2: OBSE BB1: 18.8K, OGSE BB2: 18.2K

  * Setup #3: OBSE BB1: 25.4K, OGSE BB2: 24.6K

  * Setup #4: OBSE BB1: 32.5K, OGSE BB2: 31.5K

  * Setup #5: OBSE BB1: 40.7K, OGSE BB2: 39.3K

  * Setup #6: OBSE BB1: 48.0K, OGSE BB2: 46.0K

  * Setup #7: OBSE BB1: 61.4K, OGSE BB2: 58.6K

  * Setup #8: OBSE BB1: 77.0K, OGSE BB2: 73.0K
 
 * Estimated duration of the tests for each indivitual OGSE BB setting: 2x4x4x128s + 2x15s for grating movement + 1080s OGSE BB stabilization = 1.45h

== Signal dependence on chopper frequency (1.2.8) ==
 
Line 8: Line 35:
 * Temperature settings for OGSE BB1: 16.3, 18.8, 25.4, 32.5, 40.7, 48.0, 61.4, 77.0
 * Temperature settings for OGSE BB2: 15.7, 18.2, 24.6, 31.5, 39.3, 46.0, 58.6, 73.0
 * Reset intervals: 1/16s, 1/4s, 1/2s, 2s
 * Integration capacitors: 0.1pF, 0.3pF, 1.0pF, 3.0pF
 * CUS scripts: SPEC_dynamic_range_fmilt123_OBS, SPEC_DiagHK_DynRange_block_stop, SPEC_DiagHK_DynRange_block_start
 * Measurements / Preliminary analysis : FmIltAnal/DynamicRangeSpectro
 * Pre-test setup:
   * CS1 temperature: 70.5K (fixed)
   * CS2 temperature: 76.0K (fixed)
   * Grating: 425250 (fixed)
   * Reset interval: 1/4s (fixed)
   * SPU setup: Entire slope fitting
   * Set the OGSE BB temperatures (SET_UP_OGSE_BB_TEMP_OBS, see the temperatures below)
   * Move filter wheel to "POS B" (SPEC_fltw_move)
 * CUS script: SPEC_chopper_freq_fmilt128_obs
 * Each measurement block (OGSE BB temperature setup) runs performs a tringular chopping measurement between the calibration sources and the OGSE BB1 as OGSE BB1 -- CS1 -- OGSE BB1 -- CS2 -- OBSE BB1 -- CS1 -- ...

  * Setup #1: OBSE BB1: 18.8K

  * Setup #2: OBSE BB1: 25.4K

  * Setup #3: OBSE BB1: 32.5K

  * Setup #4: OBSE BB1: 40.7K

  * Setup #5: OBSE BB1: 48.0K

  * Setup #6: OBSE BB1: 61.4K

  * Setup #7: OBSE BB1: 77.0K
 
 * Estimated duration of the tests for each indivitual OGSE BB setting: ~0.45h

Spectroscopy Setup

Dynamic range per capacitance (1.2.3)

  • Procedures : FmIltProc/DynamicRangeSpectro

  • Pre-test setup:
    • Note, that the related PTD contains a combined measurement of 1.2.3, 1.2.8 and 1.2.18 blocks, and cannot be applied in its original form.
    • Set the OGSE BB temperatures (SET_UP_OGSE_BB_TEMP_OBS, see the temperatures below)
    • Move filter wheel to "POS A" (SPEC_fltw_move)
  • CUS scripts: SPEC_dyn_range_obs, SPEC_DiagHK_DynRange_block_stop, SPEC_DiagHK_DynRange_block_start

  • Each measurement block (OGSE BB temperature setup) runs through all combinations of integration capacitor (0.1pF, 0.3pF, 1pF, 3pF) and reset interval values (1/8s, 1/4s, 1/2s, 2s), and is performed for two grating positions (GR-1 = 760000 and GR-2 = 490000).
    • Setup #1: OBSE BB1: 16.3K, OGSE BB2: 15.7K
    • Setup #2: OBSE BB1: 18.8K, OGSE BB2: 18.2K
    • Setup #3: OBSE BB1: 25.4K, OGSE BB2: 24.6K
    • Setup #4: OBSE BB1: 32.5K, OGSE BB2: 31.5K
    • Setup #5: OBSE BB1: 40.7K, OGSE BB2: 39.3K
    • Setup #6: OBSE BB1: 48.0K, OGSE BB2: 46.0K
    • Setup #7: OBSE BB1: 61.4K, OGSE BB2: 58.6K
    • Setup #8: OBSE BB1: 77.0K, OGSE BB2: 73.0K
  • Estimated duration of the tests for each indivitual OGSE BB setting: 2x4x4x128s + 2x15s for grating movement + 1080s OGSE BB stabilization = 1.45h

Signal dependence on chopper frequency (1.2.8)

  • Procedures : FmIltProc/DynamicRangeSpectro

  • Pre-test setup:
    • CS1 temperature: 70.5K (fixed)
    • CS2 temperature: 76.0K (fixed)
    • Grating: 425250 (fixed)
    • Reset interval: 1/4s (fixed)
    • SPU setup: Entire slope fitting
    • Set the OGSE BB temperatures (SET_UP_OGSE_BB_TEMP_OBS, see the temperatures below)
    • Move filter wheel to "POS B" (SPEC_fltw_move)
  • CUS script: SPEC_chopper_freq_fmilt128_obs
  • Each measurement block (OGSE BB temperature setup) runs performs a tringular chopping measurement between the calibration sources and the OGSE BB1 as OGSE BB1 -- CS1 -- OGSE BB1 -- CS2 -- OBSE BB1 -- CS1 -- ...
    • Setup #1: OBSE BB1: 18.8K
    • Setup #2: OBSE BB1: 25.4K
    • Setup #3: OBSE BB1: 32.5K
    • Setup #4: OBSE BB1: 40.7K
    • Setup #5: OBSE BB1: 48.0K
    • Setup #6: OBSE BB1: 61.4K
    • Setup #7: OBSE BB1: 77.0K
  • Estimated duration of the tests for each indivitual OGSE BB setting: ~0.45h