== Bias loops and detector heater loops== | |
Line 6: | Line 5: |
* Procedures : FmIltProc/DynamicRangeSpectro * Pre-test setup: * Note, that the related PTD contains a combined measurement of 1.2.3, 1.2.8 and 1.2.18 blocks, and cannot be applied in its original form. * Set the OGSE BB temperatures (SET_UP_OGSE_BB_TEMP_OBS, see the temperatures below) * Move filter wheel to "POS A" (SPEC_fltw_move) * CUS scripts: SPEC_dyn_range_obs, SPEC_DiagHK_DynRange_block_stop, SPEC_DiagHK_DynRange_block_start * Each measurement block (OGSE BB temperature setup) runs through all combinations of integration capacitor (0.1pF, 0.3pF, 1pF, 3pF) and reset interval values (1/8s, 1/4s, 1/2s, 2s), and is performed for two grating positions (GR-1 = 760000 and GR-2 = 490000). |
|
Line 7: | Line 14: |
* Procedures : FmIltProc/DynamicRangeSpectro | * Setup #1: OBSE BB1: 16.3K, OGSE BB2: 15.7K |
Line 9: | Line 16: |
* Setup #1: OBSE BB1: 16.3K, OGSE BB2: 15.7K | * Setup #2: OBSE BB1: 18.8K, OGSE BB2: 18.2K |
Line 11: | Line 18: |
|| || 0.1pF || 0.3pF || 1pF || 3pF || || 1/16s || || || || || 1/4s || || || || || 1/2s || || || || || 2s || || || || |
* Setup #3: OBSE BB1: 25.4K, OGSE BB2: 24.6K |
Line 17: | Line 20: |
* Setup #4: OBSE BB1: 32.5K, OGSE BB2: 31.5K | |
Line 18: | Line 22: |
|| .... || 0.1pF || 0.3pF || 1pF || 3pF || || 1/16s || || || || || PACS_Phot_Fil_Diaghk_Setup_OBS || || || || || PACS_Phot_Fil_Parameter``Loop_OBS || || || || || PACS_Diaghk_Reset_OBS || || || || || PACS_SAFE_MODE.tcl || || || || |
* Setup #5: OBSE BB1: 40.7K, OGSE BB2: 39.3K |
Line 25: | Line 24: |
* Setup #6: OBSE BB1: 48.0K, OGSE BB2: 46.0K | |
Line 26: | Line 26: |
* Setup #7: OBSE BB1: 61.4K, OGSE BB2: 58.6K | |
Line 27: | Line 28: |
* Reset intervals: 1/16s, 1/4s, 1/2s, 2s * Integration capacitors: 0.1pF, 0.3pF, 1.0pF, 3.0pF * CUS scripts: SPEC_dynamic_range_fmilt123_OBS, SPEC_DiagHK_DynRange_block_stop, SPEC_DiagHK_DynRange_block_start * Measurements / Preliminary analysis : FmIltAnal/DynamicRangeSpectro |
* Setup #8: OBSE BB1: 77.0K, OGSE BB2: 73.0K * Estimated duration of the tests for each indivitual OGSE BB setting: 2x4x4x128s + 2x15s for grating movement + 1080s OGSE BB stabilization = 1.45h |
=Spectroscopy Setup=
Dynamic range per capacitance
Procedures : FmIltProc/DynamicRangeSpectro
- Pre-test setup:
- Note, that the related PTD contains a combined measurement of 1.2.3, 1.2.8 and 1.2.18 blocks, and cannot be applied in its original form.
- Set the OGSE BB temperatures (SET_UP_OGSE_BB_TEMP_OBS, see the temperatures below)
- Move filter wheel to "POS A" (SPEC_fltw_move)
CUS scripts: SPEC_dyn_range_obs, SPEC_DiagHK_DynRange_block_stop, SPEC_DiagHK_DynRange_block_start
- Each measurement block (OGSE BB temperature setup) runs through all combinations of integration capacitor (0.1pF, 0.3pF, 1pF, 3pF) and reset interval values (1/8s, 1/4s, 1/2s, 2s), and is performed for two grating positions (GR-1 = 760000 and GR-2 = 490000).
- Setup #1: OBSE BB1: 16.3K, OGSE BB2: 15.7K
- Setup #2: OBSE BB1: 18.8K, OGSE BB2: 18.2K
- Setup #3: OBSE BB1: 25.4K, OGSE BB2: 24.6K
- Setup #4: OBSE BB1: 32.5K, OGSE BB2: 31.5K
- Setup #5: OBSE BB1: 40.7K, OGSE BB2: 39.3K
- Setup #6: OBSE BB1: 48.0K, OGSE BB2: 46.0K
- Setup #7: OBSE BB1: 61.4K, OGSE BB2: 58.6K
- Setup #8: OBSE BB1: 77.0K, OGSE BB2: 73.0K
- Estimated duration of the tests for each indivitual OGSE BB setting: 2x4x4x128s + 2x15s for grating movement + 1080s OGSE BB stabilization = 1.45h