1. Preconditions
- PACS in Spectroscopy mode
2. Procedure
Generic CUS script: PACS_Spec_Heat_Char_GeGa_Setup_OBS
- Variation of setup spectroscopy without mechanisms is executed
- SPEC_spu_Setup_OBS
- PACS_Spec_Heat_SWON_OBS
- Is the blue setting still the same ?
- PACS_Spec_Heat_Setup_OBS [310,0]
- delay 600
- PACS_Spec_Heat_Setup_OBS [0,0]
- delay 600
- Does the blue influence the red / Is anything changed in the thermal behaviour ?
- PACS_Spec_Heat_Setup_OBS [1526,0]
- delay 600
- PACS_Spec_Heat_Setup_OBS [0,0]
- delay 600
Compare Blue & Red at low current
- PACS_Spec_Heat_Setup_OBS [310,310]
- delay 600
- PACS_Spec_Heat_Setup_OBS [0,0]
- delay 600
- Explore Red at high current
- PACS_Spec_Heat_Setup_OBS [1079,1079]
- delay 600
- PACS_Spec_Heat_Setup_OBS [0,0]
- delay 600
- Repeat with PACS_Spec_Heat_Setup_OBS [2048,2048] if needed to reach curing temperature
- Full characterisation : Loop over small currents for both heaters together
- PACS_Spec_Heat_PFT_Tlow_2