Preliminary analysis disconnecting redundant harness filter wheel
Data
- Photometer filter wheel, redundant harness disconnected, treshold=2000
- 20070307/FILT_FWPHOT_test_disconn_20070307_01.tm
- Photometer filter wheel, redundant harness disconnected, treshold=4000
- 20070307/FILT_FWPHOT_test_disconn_20070307_02.tm
- Photometer filter wheel, short-circuit connector on redundant harness, treshold=2000
- 20070307/FILT_FWPHOT_test_disconn_short_20070307_01.tm
- Photometer filter wheel, short-circuit connector on redundant harness, treshold=4000
- 20070307/FILT_FWPHOT_test_disconn_short_20070307_02.tm
- Spectrometer filter wheel, redundant harness disconnected, treshold=2000
- 20070307/FILT_FWSPEC_test_disconn_20070307_01.tm
- Spectrometer filter wheel, short-circuit connector on redundant harness, treshold=2000
- 20070307/FILT_FWSPEC_test_disconn_short_20070307_02.tm
Preliminary analysis by PR
- Disconnecting the redundant chain does not improve the situation at
- all (if we really had to work like that, I would recommand to try a threshold of 6000).
- short circuit of the connectors brings us back to a "smooth" situation:
- smooth movement, and apparently much better reproducibility of stop positions.