Line 92: | Line 92: |
* mode '''OBCP_????_OBS''' calling blocks/procedures: * procedure '''SPEC_spu_setup''' * procedure '''SPEC_fltw_move''' * block '''OBCP_???''' * procedure '''SPEC_fltw_move''' * block '''OBCP_???''' * procedure '''SPEC_fltw_move''' * procedure '''SPEC_spu_reset''' |
* mode '''OBCP_chop_scan_spec_OBS''' calling blocks/procedures: * procedure '''SPEC_spu_setup''' * procedure '''SPEC_fltw_move''' * block '''OBCP_chop_scan_spec''' * procedure '''SPEC_fltw_move''' * block '''OBCP_chop_scan_spec''' * procedure '''SPEC_fltw_move''' * procedure '''SPEC_spu_reset''' * mode '''OBCP_grat_scan_nochop_OBS''' calling blocks/procedures: * procedure '''SPEC_spu_setup''' * block '''OBCP_grat_scan_nochop''' * procedure '''SPEC_spu_reset''' * mode '''OBCP_spec_2_3_chop_OBS''' calling blocks/procedures: * procedure '''SPEC_spu_setup''' * block '''OBCP_spec_2_3_chop''' * procedure '''SPEC_spu_reset''' * mode '''OBCP_grat_line_scan_chop_2_OBS''' calling blocks/procedures: * procedure '''SPEC_spu_setup''' * block '''OBCP_grat_line_scan_chop_2''' * procedure '''SPEC_spu_reset''' * mode '''OBCP_gra_scan_chop2_fast_OBS''' calling blocks/procedures: * procedure '''SPEC_spu_setup''' * block '''OBCP_gra_scan_chop2_fast''' * procedure '''SPEC_spu_reset''' * mode '''OBCP_spec_freq_switch_OBS''' calling blocks/procedures: * procedure '''SPEC_spu_setup''' * block '''OBCP_spec_freq_switch''' * procedure '''SPEC_spu_reset''' * mode '''OBCP_chop_grat_scan_cal_OBS''' calling blocks/procedures: * procedure '''SPEC_spu_setup''' * block '''OBCP_chop_grat_scan_cal''' * procedure '''SPEC_spu_reset''' |
Line 105: | Line 134: |
* block '''OBCP_???''' * procedure '''SPEC_fltw_move''' * block '''OBCP_???''' * procedure '''SPEC_fltw_move''' * procedure '''SPEC_spu_reset''' __Note__: all these modes can be started with default parameters via the generic tcl-script; the three key-OBCPs can also be started with freely selectable parameters via specific tcl-scripts |
* block '''OBCP_spec_2_3_chop''' * block '''OBCP_spec_2_3_chop_dither''' * block '''OBCP_grat_line_scan_chop_2''' * block '''OBCP_grat_scan_chop2_fast''' * block '''OBCP_grat_scan_nochop''' * block '''OBCP_spec_2_3_chop''' * block '''OBCP_chop_grat_scan_cal''' * block '''OBCP_chop_scan_spec''' * block '''OBCP_spec_freq_switch''' * procedure '''SPEC_fltw_move''' * block '''OBCP_spec_2_3_chop''' * block '''OBCP_spec_2_3_chop_dither''' * block '''OBCP_grat_line_scan_chop_2''' * block '''OBCP_grat_scan_chop2_fast''' * block '''OBCP_grat_scan_nochop''' * block '''OBCP_spec_2_3_chop''' * block '''OBCP_chop_grat_scan_cal''' * block '''OBCP_chop_scan_spec''' * block '''OBCP_spec_freq_switch''' * procedure '''SPEC_fltw_move''' * procedure '''SPEC_spu_reset''' __Note__: all these modes can be started with default parameters via the generic tcl-script; the key-OBCPs can also be started with freely selectable parameters via specific tcl-scripts |
Line 138: | Line 183: |
* block '''OBCP_???''' | * block '''OBCP_spec_2_3_chop''' |
Line 141: | Line 186: |
* block '''OBCP_???''' * block '''OBCP_???''' |
* block '''OBCP_spec_2_3_chop''' * block '''OBCP_spec_2_3_chop''' |
PACS spectrometer operation from the commanding point of view
T. Mueller, D. Cesarsky; 10-Nov-2006
Index TableOfContents(2)
1. Short Functional Tests (Ge:Ga-detector related)
1.1. SFT Warm: Heater, flasher and Ge:Ga detector chain
PACS_Spec_OBS
1.2. SFT at HeI: Heater, flasher and Ge:Ga detector chain
PACS_Spec_OBS
1.3. SFT at HeII: Heater, flasher and Ge:Ga detector chain
PACS_Spec_OBS
2. Nominal Operations
2.1. Instrument switch-on and connection of the warm electronic boxes
currently done with tcl-script PACS_Switch_On.tcl (maintained by FGB)
after execution: PACS is in SAFE mode
2.2. General Setup for Spectrometer Operation
starting out of SAFE mode
mode SPEC_setup_OBS calling the block SPEC_setup, calling the following CUS procedures:
- procedure SPEC_dpu_dmc_setup
- procedure SPEC_init_det_setup
- procedure SPEC_cre_selection
- procedure SPEC_gega_setup
- procedure SPEC_dmc_setup
- procedure SPEC_fltw_conf
- procedure SPEC_fltw_move
- procedure CONF_chopper
- procedure CHOP_on_off
- procedure CONF_cs
- procedure CS_on_off
- procedure CS_temp
- procedure CONF_grating
- procedure GRAT_on_off
- procedure SPEC_spu_setup
ending up in Spectroscopy mode, but SPU processing has not yet started
- filter is in position "A",
- chopper enabled and at position "0"
- grating is homed (with 1deg/sec) and enabled for nominal operations with a speed of 4deg/sec
- calibration sources are set to default values, but need 40-50 min stabilisation time!!!
2.3. Operation of the spectrometers (starting out of '''Spectroscopy mode''')
2.3.1. start/stop of bolometer data flow
mode SPEC_spu_setup_OBS calling procedure SPEC_spu_setup ---> spectrometer data flow running
mode SPEC_spu_reset_OBS calling procedure SPEC_spu_reset ---> spectrometer data flow stopped
2.3.2. Take data with manual chopper, grating & filter movements
mode SPEC_spu_setup_OBS calling procedure SPEC_spu_setup ---> spectrometer data flow running
mode PACS_fltw_move_OBS calling procedure SPEC_fltw_move (or PHOT_fltw_move)
or command Pacs_DMC_MOVE_CHOP_ABS
or command Pacs_DMC_MOVE_CHOP_REL
or command Pacs_DMC_MOVE_GRAT_ABS
or command Pacs_DMC_MOVE_GRAT_REL
mode SPEC_spu_reset_OBS calling procedure SPEC_spu_reset ---> spectrometer data flow stopped
Note: PACS_fltw_move_OBS can be called via a specific tcl-script where the filter has to be specified.
2.3.3. Take data with pre-defined OBCPs
mode OBCP_chop_scan_spec_OBS calling blocks/procedures:
procedure SPEC_spu_setup
procedure SPEC_fltw_move
block OBCP_chop_scan_spec
procedure SPEC_fltw_move
block OBCP_chop_scan_spec
procedure SPEC_fltw_move
procedure SPEC_spu_reset
mode OBCP_grat_scan_nochop_OBS calling blocks/procedures:
procedure SPEC_spu_setup
block OBCP_grat_scan_nochop
procedure SPEC_spu_reset
mode OBCP_spec_2_3_chop_OBS calling blocks/procedures:
procedure SPEC_spu_setup
block OBCP_spec_2_3_chop
procedure SPEC_spu_reset
mode OBCP_grat_line_scan_chop_2_OBS calling blocks/procedures:
procedure SPEC_spu_setup
block OBCP_grat_line_scan_chop_2
procedure SPEC_spu_reset
mode OBCP_gra_scan_chop2_fast_OBS calling blocks/procedures:
procedure SPEC_spu_setup
block OBCP_gra_scan_chop2_fast
procedure SPEC_spu_reset
mode OBCP_spec_freq_switch_OBS calling blocks/procedures:
procedure SPEC_spu_setup
block OBCP_spec_freq_switch
procedure SPEC_spu_reset
mode OBCP_chop_grat_scan_cal_OBS calling blocks/procedures:
procedure SPEC_spu_setup
block OBCP_chop_grat_scan_cal
procedure SPEC_spu_reset
mode SPEC_all_aots_OBS calling the following blocks/procedures:
procedure SPEC_spu_setup
procedure SPEC_fltw_move
block OBCP_spec_2_3_chop
block OBCP_spec_2_3_chop_dither
block OBCP_grat_line_scan_chop_2
block OBCP_grat_scan_chop2_fast
block OBCP_grat_scan_nochop
block OBCP_spec_2_3_chop
block OBCP_chop_grat_scan_cal
block OBCP_chop_scan_spec
block OBCP_spec_freq_switch
procedure SPEC_fltw_move
block OBCP_spec_2_3_chop
block OBCP_spec_2_3_chop_dither
block OBCP_grat_line_scan_chop_2
block OBCP_grat_scan_chop2_fast
block OBCP_grat_scan_nochop
block OBCP_spec_2_3_chop
block OBCP_chop_grat_scan_cal
block OBCP_chop_scan_spec
block OBCP_spec_freq_switch
procedure SPEC_fltw_move
procedure SPEC_spu_reset
Note: all these modes can be started with default parameters via the generic tcl-script; the key-OBCPs can also be started with freely selectable parameters via specific tcl-scripts
2.3.4. Execute AOT observations
CUS scripts (to be finalised):BR
2.4. Reset after spectrometer operation
PACS is in SPECTROSCOPY
execute mode SAFE mode
2.5. Instrument switch-off
currently done with the tcl-script PACS_Switch_Off.tcl (maintained by FGB)
- after execution: PACS is switched off
3. Specific Spectrometer Tests
3.1. Thermal Behaviour Test
starting out of SAFE mode
mode SPEC_thermal_OBS calling the following blocks/procedures:
procedure SPEC_dpu_dmc_setup
procedure CONF_chopper
procedure CHOP_on_off
block OBCP_spec_2_3_chop
procedure CONF_grating
procedure GRAT_on_off
block OBCP_spec_2_3_chop
block OBCP_spec_2_3_chop
procedure SPEC_fltw_conf
procedure SPEC_fltw_move
procedure CONF_cs
procedure CS_on_off
procedure CS_temp
procedure SPEC_dpu_dmc_reset
ending up in SAFE mode
3.2. SPU Modes & Detector Selection Tests
3.3. Other Tests
4. NOTES
- the SFT tests are described in PACS-ME-TP-007
some specific spectrometer tests, like the thermal behaviour measurement, start directly out of SAFE mode, while most of the tests will be executed after the SPEC_setup_OBS when the instrument is in Spectroscoy mode
- the above mentioned CUS scripts (with a few exceptions) have been tested and are available in the CUS database and from the CVS library
- the AOT related CUS-scripts are still under testing