1. PACS bolometer operation from the commanding point of view
T. Mueller, D. Cesarsky 19-May-2006
- Instrument switch-on and connection of the warm electronic boxes
- - currently done with tcl-script "PACS_Switch_On.tcl" (maintained by FGB): - after execution: PACS is in SAFE mode
- Cooler recycling (only at HeII temperatures)
- - starting out of SAFE mode - CUS script: BOLO_cool_recycle
- (no sub-procedures are called)
- General Setup for Photometer Operation
- (at warm, HeI, HeII temperatures and after cooler recycling)
- - starting out of SAFE mode - CUS script:
- block PHOT_setup calling the following CUS procedures
- (will be renamed to PHOT_orbit_prologue):
- - procedure PHOT_dpu_dmc_setup - procedure PHOT_dmc_setup
- - procedure PHOT_fltw_conf (*) - procedure CONF_chopper (*) - procedure CONF_cs (*)
- environment!
- - procedure PHOT_dpu_dmc_setup - procedure PHOT_dmc_setup
- started. Note: filter is in position "A", chopper at position "0", calibration sources are set to default values, but need 30-40 min stabilisation time!!!
- Operation of the bolometers (starting out of PHOTOMETRY mode)
- A) start/stop of bolometer data flow
- CUS scripts:
- - procedure PHOT_spu_setup
---> bolometer data flow running
- - procedure PHOT_spu_reset
Change of the bolometer data/sequence mode and gain
- CUS scripts:
- - procedure PHOT_change_data_mode - procedure PHOT_change_seq_mode - procedure PHOT_set_gain
C) Take data with manual chopper & filter movements
- CUS scripts:
- - procedure PHOT_spu_setup
---> bolometer data flow running
- - procedure PHOT_fltw_move - command Pacs_DMC_MOVE_CHOP_ABS - command Pacs_DMC_MOVE_CHOP_REL
- - procedure PHOT_spu_reset
- CUS scripts:
- - procedure PHOT_spu_setup
---> bolometer data flow running
- - block OBCP_chopped_photometry - block OBCP_chopped_photometry_dither - block OBCP_phot_freezeframe - block OBCP_phot_cal_I - block OBCP_phot_cal_II - block OBCP_phot_cal_III - block OBCP_phot_staring - block OBCP_phot_chopper_scan
- (not yet available!!!)
- - block PHOT_all_aots_OBS
- - procedure PHOT_spu_reset
- - block OBCP_chopped_photometry - block OBCP_chopped_photometry_dither - block OBCP_phot_freezeframe - block OBCP_phot_cal_I - block OBCP_phot_cal_II - block OBCP_phot_cal_III - block OBCP_phot_staring - block OBCP_phot_chopper_scan
- CUS scripts (to be finalised):
- PHOT_AOT
- PacsPhotDefaults - PacsPhotEstimatorOBCPn
- - PacsDureeOBCP5
- - DMC_phot_2_3_chop
- - DMC_staring_phot
- - composite_nodding - repeated_raster_with_hold - line_scan_with_hold
- - PHOT_aot_prologue
- - PHOT_spu_setup
- - OBCP_staring_photometry
- - OBCP_chopped_photometry_dither
- - PacsDureeOBCP5
- PHOT_AOT
- CUS scripts:
- Reset after photometer operation
- - PACS is in PHOTOMETRY or any other mode - going into SAFE mode:
- CUS scripts:
- - block OBCP_SAFE
- CUS scripts:
- Instrument switch-off
- - currently done with a dedicated tcl-script (maintained by FGB) - after execution: PACS is switched off
- NOTES
- - there are more CUS scripts related to the SFT warm, HeI and HeII
- which are not mentioned here (see PACS-ME-TP-007)
- measurements, start directly out of SAFE mode, while most of the tests will be executed after the "PHOT_setup"
- tested and are available in the CUS database and from the CVS library