Differences between revisions 18 and 19
Revision 18 as of 2006-11-09 11:33:58
Size: 11457
Comment:
Revision 19 as of 2006-11-09 12:40:59
Size: 10925
Comment:
Deletions are marked like this. Additions are marked like this.
Line 18: Line 18:
 * '''PACS_Phot_SFT_Cooler_Warm_OBS''', calling '''PACS_Phot_SFT_Cooler_Warm'''[[BR]]

 * '''PACS_Phot_DPU_DMC_Setup_OBS''', calling '''PACS_Phot_DPU_DMC_Setup'''[[BR]]
 * '''PACS_Phot_Switchon_OBS''', calling '''PACS_Phot_Switchon'''[[BR]]
 * '''PACS_Phot_Sequencer_Setup_OBS''', calling '''PACS_Phot_Sequencer_Setup'''[[BR]]
 * '''PACS_Phot_SPU_Setup_OBS''', calling '''PACS_Phot_SPU_Setup'''[[BR]]
 * '''PACS_Phot_Warm_Startinputsignal_OBS''', calling '''PACS_Phot_Warm_Startinputsignal''' (Duration: 637 sec)[[BR]]
 *
'''PACS_Phot_SPU_Reset_OBS''', calling '''PACS_Phot_SPU_Reset'''[[BR]]
 * '''PACS_Phot_Switchoff_OBS''', calling '''PACS_Phot_Switchoff'''[[BR]]
 * '''PACS_Phot_SFT_Cooler_Warm_OBS''', calling '''PACS_Phot_SFT_Cooler_Warm'''

 * '''PACS_Phot_DPU_DMC_Setup_OBS''', calling '''PACS_Phot_DPU_DMC_Setup'''
 * '''PACS_Phot_Switchon_OBS''', calling '''PACS_Phot_Switchon'''
 * '''PACS_Phot_Sequencer_Setup_OBS''', calling '''PACS_Phot_Sequencer_Setup'''
 * '''PACS_Phot_SPU_Setup_OBS''', calling '''PACS_Phot_SPU_Setup'''
 * '''PACS_Phot_Warm_Startinputsignal_OBS''', calling '''PACS_Phot_Warm_Startinputsignal''' (Duration: 637 sec)
 * '''PACS_Phot_SPU_
Reset_OBS''', calling '''PACS_Phot_SPU_Reset'''
 * '''PACS_Phot_Switchoff_OBS''', calling '''PACS_Phot_Switchoff'''
Line 33: Line 33:
 * '''PACS_Phot_SFT_Cooler_OBS''', calling '''PACS_Phot_SFT_Cooler'''[[BR]]

 * '''PACS_Phot_DPU_DMC_Setup_OBS''', calling '''PACS_Phot_DPU_DMC_Setup'''[[BR]]
 * '''PACS_Phot_Switchon_OBS''', calling '''PACS_Phot_Switchon'''[[BR]]
 * '''PACS_Phot_Sequencer_Setup_OBS''', calling '''PACS_Phot_Sequencer_Setup'''[[BR]]
 * '''PACS_Phot_SPU_Setup_OBS''', calling '''PACS_Phot_SPU_Setup'''[[BR]]
 * '''PACS_Phot_SFT_Cooler_OBS''', calling '''PACS_Phot_SFT_Cooler'''

 * '''PACS_Phot_DPU_DMC_Setup_OBS''', calling '''PACS_Phot_DPU_DMC_Setup'''
 * '''PACS_Phot_Switchon_OBS''', calling '''PACS_Phot_Switchon'''
 * '''PACS_Phot_Sequencer_Setup_OBS''', calling '''PACS_Phot_Sequencer_Setup'''
 * '''PACS_Phot_SPU_Setup_OBS''', calling '''PACS_Phot_SPU_Setup'''
 * '''PACS_Phot_Cold_Startinputsignal_OBS''', calling '''PACS_Phot_Cold_Startinputsignal''' (Duration: 637 sec)
 * '''PACS_Phot_SPU_Reset_OBS''', calling '''PACS_Phot_SPU_Reset'''
 * '''PACS_Phot_Switchoff_OBS''', calling '''PACS_Phot_Switchoff'''

__Note__: No specific autonomy functions are involved here (except AF#6 to check for the communication between DMC and BOLC)


== SFT at HeII (cooler not recycled): Sorption Cooler and Bolometer detector chain ==

 * '''PACS_Phot_SFT_Cooler_OBS''', calling '''PACS_Phot_SFT_Cooler'''

 * '''PACS_Phot_DPU_DMC_Setup_OBS''', calling '''PACS_Phot_DPU_DMC_Setup'''
 * '''PACS_Phot_Switchon_OBS''', calling '''PACS_Phot_Switchon'''
 * '''PACS_Phot_Sequencer_Setup_OBS''', calling '''PACS_Phot_Sequencer_Setup'''
 * '''PACS_Phot_SPU_Setup_OBS''', calling '''PACS_Phot_SPU_Setup'''
Line 40: Line 55:
 * '''PACS_Phot_SPU_Reset_OBS''', calling '''PACS_Phot_SPU_Reset'''[[BR]]
 * '''PACS_Phot_Switchoff_OBS''', calling '''PACS_Phot_Switchoff'''[[BR]]
 * '''PACS_Phot_SPU_Reset_OBS''', calling '''PACS_Phot_SPU_Reset'''
 * '''PACS_Phot_Switchoff_OBS''', calling '''PACS_Phot_Switchoff'''
Line 45: Line 60:

== SFT at HeII (cooler not recycled): Sorption Cooler and Bolometer detector chain ==

 * '''PACS_Phot_SFT_Cooler_OBS''', calling '''PACS_Phot_SFT_Cooler'''[[BR]]

 * '''PACS_Phot_DPU_DMC_Setup_OBS''', calling '''PACS_Phot_DPU_DMC_Setup'''[[BR]]
 * '''PACS_Phot_Switchon_OBS''', calling '''PACS_Phot_Switchon'''[[BR]]
 * '''PACS_Phot_Sequencer_Setup_OBS''', calling '''PACS_Phot_Sequencer_Setup'''[[BR]]
 * '''PACS_Phot_SPU_Setup_OBS''', calling '''PACS_Phot_SPU_Setup'''[[BR]]
 * '''PACS_Phot_Cold_Startinputsignal_OBS''', calling '''PACS_Phot_Cold_Startinputsignal''' (Duration: 637 sec)[[BR]]
 * '''PACS_Phot_SPU_Reset_OBS''', calling '''PACS_Phot_SPU_Reset'''[[BR]]
 * '''PACS_Phot_Switchoff_OBS''', calling '''PACS_Phot_Switchoff'''[[BR]]

__Note__: No specific autonomy functions are involved here (except AF#6 to check for the communication between DMC and BOLC)
Line 67: Line 67:
 * currently done with tcl-script '''PACS_Switch_On.tcl''' (maintained by FGB)[[BR]]
 * after execution: PACS is in SAFE mode, the bolometer groups are off, but the temperature sensors are on[[BR]]
 * currently done with tcl-script '''PACS_Switch_On.tcl''' (maintained by FGB)
 * after execution: PACS is in SAFE mode, the bolometer groups are off, but the temperature sensors are on
Line 76: Line 76:
 * starting out of SAFE mode (groups off, temperature sensors on) [[BR]]
 * mode '''BOLO_cooler_OBS''' calling the CUS procedure '''BOLO_cool_recycle''':[[BR]]
             - no sub-procedures are called, the groups and temperature sensors are not touched[[BR]]
             - HK_list is set to PHOT_HK and reset to "NON-PRIME" in the end[[BR]]
             - autonomy function 14 is enabled, Duration: 7928 sec[[BR]]
 * ending up in SAFE mode (groups are off, temperature sensors on, HSP heater current at 1.18 mA)[[BR]]
 * starting out of SAFE mode (groups off, temperature sensors on)
 * mode '''BOLO_cooler_OBS''' calling the CUS procedure '''BOLO_cool_recycle''':
   * no sub-procedures are called, the groups and temperature sensors are not touched
   * HK_list is set to PHOT_HK and reset to "NON-PRIME" in the end
   * autonomy function 14 is enabled, Duration: 7928 sec
 * ending up in SAFE mode (groups are off, temperature sensors on, HSP heater current at 1.18 mA)
Line 90: Line 90:
 * mode '''PHOT_setup_OBS''' calling the block '''PHOT_setup''', calling the following CUS procedures:[[BR]]
               -
procedure PHOT_dpu_dmc_setup[[BR]]
               -
procedure PHOT_dmc_setup[[BR]]
                              -
procedure PHOT_fltw_conf [[BR]]
                              -
procedure CONF_chopper [[BR]]
                              -
procedure CONF_cs [[BR]]
               -
procedure PHOT_switchon_groups[[BR]]
               -
procedure PHOT_set_bias (*)[[BR]]
               -
procedure PHOT_seq_setup[[BR]]
               -
procedure PHOT_change_data_mode[[BR]]
               -
procedure PHOT_change_seq_mode[[BR]]
               -
procedure PHOT_set_gain(**)[[BR]]
               - procedure PHOT_spu_setup[[BR]]

    (*) the bias tables to be loaded might change![[BR]]
 * mode '''PHOT_setup_OBS''' calling the block '''PHOT_setup''', calling the following CUS procedures:
   *
procedure PHOT_dpu_dmc_setup
   *
procedure PHOT_dmc_setup
     *
procedure PHOT_fltw_conf
     *
procedure CONF_chopper
     *
procedure CONF_cs
   *
procedure PHOT_switchon_groups
   *
procedure PHOT_set_bias (*)
   *
procedure PHOT_seq_setup
   *
procedure PHOT_change_data_mode
   *
procedure PHOT_change_seq_mode
   *
procedure PHOT_set_gain(**)
   * procedure PHOT_spu_setup

    (*) the bias tables to be loaded might change!
Line 117: Line 117:
        mode '''PHOT_spu_setup_OBS''' calling procedure '''PHOT_spu_setup''' ---> bolometer data flow running [[BR]]
        mode '''PHOT_spu_reset_OBS''' calling procedure '''PHOT_spu_reset''' ---> bolometer data flow stopped [[BR]]
 * mode '''PHOT_spu_setup_OBS''' calling procedure '''PHOT_spu_setup''' ---> bolometer data flow running
 * mode '''PHOT_spu_reset_OBS''' calling procedure '''PHOT_spu_reset''' ---> bolometer data flow stopped
Line 122: Line 122:
        - procedure '''PHOT_change_data_mode'''[[BR]]
        - procedure '''PHOT_change_seq_mode'''[[BR]]
        - procedure '''PHOT_set_gain'''[[BR]]

        - or via MSTK with individual commands
   * procedure '''PHOT_change_data_mode'''[[BR]]
   * procedure '''PHOT_change_seq_mode'''[[BR]]
   * procedure '''PHOT_set_gain'''[[BR]]

   * or via MSTK with individual commands

Index TableOfContents(2)

PACS bolometer operation from the commanding point of view


T. Mueller, D. Cesarsky; 08-Nov-2006


1. Short functional tests


1.1. SFT Warm: Sorption Cooler and Bolometer detector chain

  • PACS_Phot_SFT_Cooler_Warm_OBS, calling PACS_Phot_SFT_Cooler_Warm

  • PACS_Phot_DPU_DMC_Setup_OBS, calling PACS_Phot_DPU_DMC_Setup

  • PACS_Phot_Switchon_OBS, calling PACS_Phot_Switchon

  • PACS_Phot_Sequencer_Setup_OBS, calling PACS_Phot_Sequencer_Setup

  • PACS_Phot_SPU_Setup_OBS, calling PACS_Phot_SPU_Setup

  • PACS_Phot_Warm_Startinputsignal_OBS, calling PACS_Phot_Warm_Startinputsignal (Duration: 637 sec)

  • PACS_Phot_SPU_Reset_OBS, calling PACS_Phot_SPU_Reset

  • PACS_Phot_Switchoff_OBS, calling PACS_Phot_Switchoff

Note: No specific autonomy functions are involved here (except AF#6 to check for the communication between DMC and BOLC)

1.2. SFT at HeI: Sorption Cooler and Bolometer detector chain

  • PACS_Phot_SFT_Cooler_OBS, calling PACS_Phot_SFT_Cooler

  • PACS_Phot_DPU_DMC_Setup_OBS, calling PACS_Phot_DPU_DMC_Setup

  • PACS_Phot_Switchon_OBS, calling PACS_Phot_Switchon

  • PACS_Phot_Sequencer_Setup_OBS, calling PACS_Phot_Sequencer_Setup

  • PACS_Phot_SPU_Setup_OBS, calling PACS_Phot_SPU_Setup

  • PACS_Phot_Cold_Startinputsignal_OBS, calling PACS_Phot_Cold_Startinputsignal (Duration: 637 sec)

  • PACS_Phot_SPU_Reset_OBS, calling PACS_Phot_SPU_Reset

  • PACS_Phot_Switchoff_OBS, calling PACS_Phot_Switchoff

Note: No specific autonomy functions are involved here (except AF#6 to check for the communication between DMC and BOLC)

1.3. SFT at HeII (cooler not recycled): Sorption Cooler and Bolometer detector chain

  • PACS_Phot_SFT_Cooler_OBS, calling PACS_Phot_SFT_Cooler

  • PACS_Phot_DPU_DMC_Setup_OBS, calling PACS_Phot_DPU_DMC_Setup

  • PACS_Phot_Switchon_OBS, calling PACS_Phot_Switchon

  • PACS_Phot_Sequencer_Setup_OBS, calling PACS_Phot_Sequencer_Setup

  • PACS_Phot_SPU_Setup_OBS, calling PACS_Phot_SPU_Setup

  • PACS_Phot_Cold_Startinputsignal_OBS, calling PACS_Phot_Cold_Startinputsignal (Duration: 637 sec)BR

  • PACS_Phot_SPU_Reset_OBS, calling PACS_Phot_SPU_Reset

  • PACS_Phot_Switchoff_OBS, calling PACS_Phot_Switchoff

Note: No specific autonomy functions are involved here (except AF#6 to check for the communication between DMC and BOLC)


2. Nominal operations


2.1. Instrument switch-on and connection of the warm electronic boxes

  • currently done with tcl-script PACS_Switch_On.tcl (maintained by FGB)

  • after execution: PACS is in SAFE mode, the bolometer groups are off, but the temperature sensors are on

Note: Autonomy function #6 is enabled to check for the communication between DMC and BOLC

2.2. Cooler recycling (only at HeII temperatures)

  • starting out of SAFE mode (groups off, temperature sensors on)
  • mode BOLO_cooler_OBS calling the CUS procedure BOLO_cool_recycle:

    • no sub-procedures are called, the groups and temperature sensors are not touched
    • HK_list is set to PHOT_HK and reset to "NON-PRIME" in the end
    • autonomy function 14 is enabled, Duration: 7928 sec
  • ending up in SAFE mode (groups are off, temperature sensors on, HSP heater current at 1.18 mA)

Note: At the end of the recycling the autonomy function 14 is enabled. It checks for TEMP_EV, TEMP_FPU1, TEMP_FPU2: --> HK DEC SOFT is generated if one or more are out of the range [260,320] mK; --> GO SAFE is generated if one or more are above 400 mK

2.3. General Setup for Photometer Operation (after cooler recycling)

  • starting out of SAFE mode (groups off, temperature sensors on)
  • mode PHOT_setup_OBS calling the block PHOT_setup, calling the following CUS procedures:

    • procedure PHOT_dpu_dmc_setup
    • procedure PHOT_dmc_setup
      • procedure PHOT_fltw_conf
      • procedure CONF_chopper
      • procedure CONF_cs
    • procedure PHOT_switchon_groups
    • procedure PHOT_set_bias (*)
    • procedure PHOT_seq_setup
    • procedure PHOT_change_data_mode
    • procedure PHOT_change_seq_mode
    • procedure PHOT_set_gain(**)
    • procedure PHOT_spu_setup
      • (*) the bias tables to be loaded might change! (**) the gain setting can be 'high' or 'low'
  • ending up in PHOTOMETRY mode, but SPU processing has not yet started
  • filter is in position "A", chopper at position "0"
  • calibration sources are set to default values, but need 40-50 min stabilisation time!!!

Note: At the end of the PHOT_setup the autonomy function 17 "generate_event_pwr" is enabled. It checks for the HK entries: PC_PWR_ANA_P_#, PC_PWR_ANA_N_# and PC_PWR_DIG_#, with # = [1,2,3,4,5,6,7]

2.4. Operation of the bolometers (starting out of PHOTOMETRY mode)

2.4.1. start/stop of bolometer data flow

  • mode PHOT_spu_setup_OBS calling procedure PHOT_spu_setup ---> bolometer data flow running

  • mode PHOT_spu_reset_OBS calling procedure PHOT_spu_reset ---> bolometer data flow stopped

2.4.2. Change of the bolometer data/sequence mode and gain

  • procedure PHOT_change_data_modeBR

  • procedure PHOT_change_seq_modeBR

  • procedure PHOT_set_gainBR

  • or via MSTK with individual commands

2.4.3. Take data with manual chopper & filter movements

  • mode PHOT_spu_setup_OBS calling procedure PHOT_spu_setup ---> bolometer data flow runningBR

    • - block PHOT_fltw_move_OBS calling procedure PHOT_fltw_moveBR - command Pacs_DMC_MOVE_CHOP_ABSBR - command Pacs_DMC_MOVE_CHOP_RELBR

    mode PHOT_spu_reset_OBS calling procedure PHOT_spu_reset ---> bolometer data flow stoppedBR

2.4.4. Take data with pre-defined OBCPs

  • mode PHOT_spu_setup_OBS calling procedure PHOT_spu_setup ---> bolometer data flow running BR

    • - block OBCP_chopped_photometry_OBSBR - block OBCP_phot_chopper_scan_OBSBR - block OBCP_fixed_fixed_photo_OBSBR

    mode PHOT_spu_reset_OBS calling procedure PHOT_spu_reset ---> bolometer data flow stopped BR or

    mode PHOT_spu_setup_OBS calling procedure PHOT_spu_setup ---> bolometer data flow running BR

    • - mode PHOT_all_aots_OBS calling the following procedures:BR

      • - procedure OBCP_chopped_photometryBR - procedure OBCP_chopped_photometry_ditherBR - procedure OBCP_phot_freezeframeBR - procedure OBCP_phot_cal_IBR - procedure OBCP_phot_cal_IIBR - procedure OBCP_phot_cal_IIIBR - procedure OBCP_phot_staringBR - procedure OBCP_phot_chopper_scanBR - procedure OBCP_fixed_fixed_photoBR

    mode PHOT_spu_reset_OBS calling procedure PHOT_spu_reset ---> bolometer data flow stopped BR

Note: all OBCPs can be started with default parameters via the generic tcl-script; the key-OBCPs can also be started with freely selectable parameters via specific tcl-scripts

2.4.5. Execute AOT observations

  • CUS scripts (to be finalised):BR

    • PHOT_AOTBR

      • - PacsPhotDefaultsBR - PacsPhotEstimatorOBCPnBR

        • - PacsDureeOBCP5BR

          • - DMC_phot_2_3_chopBR

          - PacsDureeOBCP7BR

          • - DMC_staring_photBR

        Logic chooses and issues a pointing mode out of:BR

        • - composite_noddingBR - repeated_raster_with_holdBR - line_scan_with_holdBR

        - WriteOBSIDBR - PacsPhotSlewCalBR

        • - PHOT_aot_prologueBR

          • - PHOT_spu_setupBR

          - PHOT_fltw_moveBR - OBCP_chopped_photometryBR

        - PacsPhotCommandOBCP7BR

        • - OBCP_staring_photometryBR

        - PacsPhotCommandOBCP5BR

        • - OBCP_chopped_photometry_ditherBR

        - PHOT_aot_epilogueBR - WriteEndIDBR - PacsPhotNoiseRMSBR - PacsMessageCenterBR

Note: the default OBCP for point-source photometry will probably be the OBCP_fixed_fixed_photo.

2.5. Reset after photometer operation

  • PACS is in PHOTOMETRY or any other mode
  • execute mode PHOT_switchoff_groups_OBS calling procedure PHOT_switchoff_groups (this will reset the biases sequentially, disable the AF 17, and switch off the groups)

  • going into SAFE mode (this will disable the AF 14 for the FPU temperature checks)
  • after execution: PACS is in SAFE mode (groups are off, temperature sensors on, AF 6 enabled to check for the communication between DMC and BOLC)

2.6. Instrument switch-off

  • currently done with the tcl-script PACS_Switch_Off.tcl (maintained by FGB) (this will disable the AF 6 for the DMC-BOLC communication check)

  • after execution: PACS is switched off


3. NOTES


  • the SFT tests are described in PACS-ME-TP-007
  • some specific photometer tests, like the thermal behaviour measurements, start directly out of SAFE mode, while most of the tests will be executed after the PHOT_setup

  • the above mentioned CUS scripts (with a few exceptions) have been tested and are available in the CUS database and from the CVS library
  • the AOT related CUS-scripts are still under testing

Herschel: PACS/FM_ILT_Procedures/PhotOperations (last edited 2009-07-15 14:32:37 by localhost)