= Chopper Control Loop Parameters (CALU table) =
------------------------------------------------------------------------------------

 * the valid control loop parameters are defined in the file '''CONFChopper''' located on pacs5 in the directory '''toREG/CALU'''
= Chopper Control Loop Parameters (CALU table) - Version: 06 March 2007 =
------------------------------------------------------------------------------------

 * the valid control loop parameters are defined in the file '''CONFChopper''' located on '''oper@pacs5''' in the directory '''toREG/CALU'''
Line 51: Line 51:
 * current default: '''CSLopt3-600'''
Line 53: Line 54:
   * ColdOPdef
     * '''working but not useful'''
     * defined on Feb 22, 2007 (MN, HD)
     * taken from CSL EGSE measurement (Alain)
     * used for MPE EGSE tests
     * assumed DMC QM amplification factor
     * Rate: 29 units, for chopper pos +/- 12000 units at 100 ms ramp
     * OutputLimit: +/- 24 mA (6000 units), maximum allowed chopper current
     * PosLimit: +/- 13000 units, maximum allowed chopper deflection
     * notch filter set to 480 Hz with DA=2

   * ColdOPzero
     * '''working but not useful'''
     * defined on Feb 22, 2007 (MN, HD)
     * same settings as column ColdOPdef
     * mechanical ZP offset register set to zero

   * ZEISSzero
     * '''used for testing only'''
     * defined on Feb 23, 2007 (MN, HD)
     * ZEISS default parameters translated to CSL units
     * CSL DMC FM amplification factor of 50.09 assumed
     * notch filter set to 900 Hz with DA=2
     * mechanical ZP offset register set to zero

   * ZEISSoffset
     * '''used for testing only'''
     * defined on Feb 23, 2007 (MN, HD)
     * same settings as column ZEISSzero
     * mechanical offset register set to -1195 as measured with ZEISSzero

   * ZEISSoffset2
     * '''to be preferred to ZEISSoffset'''
     * defined on Feb 23, 2007 (MN, HD)
     * same settings as column ZEISSoffset
     * notch filter set to 480 Hz with DA=2

   * CSLopt1
     * '''can be used as a fallback parameter set (stable)'''
     * defined on Feb 28, 2007 (MN)
     * settings determined with CSL EGSE at MPE (Benoit)
     * parameter set optimised for chopper deflection +/- 4 deg
     * '''WARNING! Oscillations at deflections > 4 deg (+/- 12000 units)!'''
     * OutputLimit: +/- 65 mA (16013 units)
     * PosLimit: +/- 14000 units
     * PosOffset: -1195 units
     * Rate: 310 units for chopper pos +/- 26000 units at 10 ms ramp
     * notch filter set to 550 Hz with DA=3

   * CSLopt2-600
     * '''Dangerous parameters! Do not use!'''
     * defined on Mar 02, 2007 (MN)
     * settings determined with CSL EGSE at MPE (Benoit)
     * notch filter set to 600 Hz with DA=3
     * worked with CSL EGSE

   * CSLopt2-550
     * '''not tested, potentially dangerous'''
     * defined on Mar 02, 2007 (MN)
     * adapted from CSLopt2-600
     * parameter set optimised for chopper deflection +/- 4 deg and also working for +/- 9 deg
     * OutputLimit: +/- 65 mA (16013 units)
     * PosLimit: +/- 25000 units
     * PosOffset: -1195 units
     * Rate: 310 units for chopper pos +/- 26000 units at 10 ms ramp
     * notch filter set to 550 Hz with DA=3
     * effective transition times:
       * approx. 25 ms at +/- 4 deg
       * approx. 60 ms at +/- 9 deg

   * CSLopt3-600
     * '''Working well. This is the default parameter set for now.'''
     * defined on Mar 02, 2007 (MN)
     * settings determined with CSL EGSE at MPE (Benoit)
     * parameter set optimised for chopper deflection +/- 4 deg and also working for +/- 9 deg
     * OutputLimit: +/- 110 mA (27100 units)
     * PosLimit: +/- 25000 units
     * PosOffset: -1195 units
     * Rate: 550 units for chopper pos +/- 26000 units at 5 ms ramp
     * notch filter set to 600 Hz with DA=3
     * effective transition times:
       * approx. 17 ms at +/- 4 deg
       * approx. 60 ms at +/- 9 deg

   * CSLopt3-550
     * '''not tested, potentially dangerous'''
     * defined on Mar 02, 2007 (MN)
     * same as CSLopt3-600
     * notch filter set to 550 Hz with DA=3
     * PosLimit: +/- 14000 units (+/- 4 deg) for safety
|| ||''ColdOPdef'' ||''ColdOPzero'' ||''ZEISSzero'' ||''ZEISSoffset'' ||''ZEISSoffset2''||
||Remark ||working but not useful||working but not useful||for testing only ||for testing only ||to be preferred to ZEISSoffset||
||Usage ||''do not use'' ||''do not use'' ||''do not use'' ||''do not use'' ||''do not use'' ||
||Date of creation||Feb 22, 2007 (MN, HD) ||Feb 22, 2007 (MN, HD) ||Feb 23, 2007 (MN, HD)||Feb 23, 2007 (MN, HD)||Feb 23, 2007 (MN, HD)||
||Source ||CSL EGSE at MPE (Alain)||CSL EGSE at MPE (Alain)||ZEISS EICD ||ZEISS EICD ||ZEISS EICD||
||PosLimit ||+/- 4.5 deg (13000 units)||+/- 4.5 deg (13000 units)||+/- 5.5 deg (16000 units)||+/- 10.5 deg (28500 units)||+/- 10.5 deg (28500 units)||
||OutputLimit ||+/- 24 mA (6000 units)||+/- 24 mA (6000 units)||+/- 30 mA (7500 units)||+/- 45 mA (11086 units)||+/- 45 mA (11086 units)||
||Rate ||29 units, calculated for chopper pos +/- 12000 units at 100 ms ramp ||29 units, calculated for chopper pos +/- 12000 units at 100 ms ramp ||29 units, calculated for chopper pos +/- 12000 units at 100 ms ramp ||29 units, calculated for chopper pos +/- 12000 units at 100 ms ramp ||29 units, calculated for chopper pos +/- 12000 units at 100 ms ramp ||
||Notch filter ||480 Hz, DA=2 ||480 Hz, DA=2 ||900 Hz, DA=2 ||900 Hz, DA=2 ||480 Hz, DA=2||
||DMC amplification||50.243 ||50.243 ||50.09 ||50.09 ||50.09 ||
||Offset mech. ZP ||-1125 units ||0 units ||0 units ||-1195 units ||-1195 units ||
||Transition times||not measured||not measured||not measured||not measured||not measured||

|| ||'''CSLopt1'''||''CSLopt2-600''||''CSLopt2-550''||'''CSLopt3-600'''||''CSLopt3-550''||
||Remark ||'''stable fallback parameter set''', ''Warning! Oscillations occur at deflections > 4.1 deg!''||'''Dangerous! Can damage chopper!''', worked with CSL EGSE||'''potentially dangerous''', ''not tested''||Working well. This is the default parameter set for now.||'''potentially dangerous''', ''not tested''||
||Usage ||''use if other settings fail''||'''never use'''||''do not use''||'''use as default'''||''do not use''||
||Date of creation||Feb 28, 2007 (MN)||Mar 02, 2007 (MN)||Mar 02, 2007 (MN)||Mar 02, 2007 (MN)||Mar 02, 2007 (MN)||
||Source ||CSL EGSE at MPE (Benoit)||CSL EGSE at MPE (Benoit)||CSL EGSE at MPE (Benoit)||CSL EGSE at MPE (Benoit)||CSL EGSE at MPE (Benoit)||
||PosLimit ||+/- 5 deg (14000 units)||+/- 9.5 deg (25000 units)||+/- 9.5 deg (25000 units)||+/- 9.5 deg (25000 units)||+/- 5 deg (14000 units)||
||OutputLimit ||+/- 65 mA (16013 units)||+/- 65 mA (16013 units)||+/- 65 mA (16013 units)||+/- 110 mA (27100 units)||+/- 110 mA (27100 units)||
||Rate ||310 units, calculated for chopper pos +/- 26000 units at 10 ms ramp||310 units, calculated for chopper pos +/- 26000 units at 10 ms ramp||310 units, calculated for chopper pos +/- 26000 units at 10 ms ramp||550 units, calculated for chopper pos +/- 26000 units at 5 ms ramp||550 units, calculated for chopper pos +/- 26000 units at 5 ms ramp||
||Notch filter ||550 Hz, DA=3 ||600 Hz, DA=3 ||550 Hz, DA=3 ||600 Hz, DA=3 ||550 Hz, DA=3 ||
||Offset mech. ZP ||-1195 units ||-1195 units ||-1195 units ||-1195 units ||-1195 units ||
||Transition time at +/- 4 deg||slow ||approx. 25 ms||approx. 25 ms||approx. 17 ms||approx. 17 ms||
||Transition time at +/- 9 deg||slow ||approx. 60 ms||approx. 60 ms||approx. 60 ms||approx. 60 ms||

H. Dannerbauer; 07-Dec-2006

Index TableOfContents(2)


1. Determination of Chopper Offset


  • for detailed test description see PTD_2.3.1 (->file attachment:chopper.pdf) or in CVS

  • Chopper has to be switched-off and controller disabled
  • upload Chopper CPRs with position offset of zero using tcl-script CONF_Chopper_OBS.tcl with parameters fileU and calUcol

  • start PacketRecorder

  • start diag. HK with SPEC_Chopper_dhk_imt_OBS using Generic_CUS_script.tcl

  • determine chopper offset (chop_offset) from DM_CHOP_CUR_POS
  • include the new chopper offset in chopper control parameter set
  • stop diag. HK with SPEC_Chopper_dhk_stop_OBS using Generic_CUS_script.tcl

  • close PacketRecorder

  • switch-on chopper
  • load up chopper control parameters including the new offset
    • -> tcl-script CONF_chopper_OBS.tcl with parameters fileU and calUcol

  • enable chopper controller in order to upload new control parameter set
    • -> EnDis_chopper_OBS.tcl and parameter onoff ON


2. Chopper Open Loop Measurements


  • for detailed test description see PTD_2.3.1 (->file attachment:chopper.pdf) or in CVS

  • it is recommended to upload proper control parameter's using tcl-script CONF_chopper_OBS.tcl with parameters fileU and calUcol

  • test should be operated in nominal spectroscopic mode; to do this execute the GeGa SFT; e.g. for He2 use PACS_GeGa_SFT_Init_Cold_He2_OBS.cus (Generic_CUS_script.tcl)

  • chopper controller has to be at least disabled
  • start diag. HK using SPEC_Chopper_dhk_4hk_1khz_OBS.cus (Generic_CUS_script.tcl)

  • start PacketRecorder

  • start script SPEC_Chopper_OpenLoop_MoveAbs_OBS.tcl given following parameters: chop_max_ma (maximum current in mA), chop_step_ma (current step width in mA) and chop_damptime (plateau duration in sec)

  • close PacketRecorder

  • stop diag. HK using SPEC_Chopper_dhk_stop_OBS.cus (Generic_CUS_script.tcl)

  • at the end of the test chopper controller is disabled; if necessary switch-off chopper


3. Chopper Control Loop Parameters (CALU table) - Version: 06 March 2007


  • the valid control loop parameters are defined in the file CONFChopper located on oper@pacs5 in the directory toREG/CALU

  • it contains the columns accessible for the upload of chopper parameters
  • current default: CSLopt3-600

  • the individual columns are characterised as follows:

ColdOPdef

ColdOPzero

ZEISSzero

ZEISSoffset

ZEISSoffset2

Remark

working but not useful

working but not useful

for testing only

for testing only

to be preferred to ZEISSoffset

Usage

do not use

do not use

do not use

do not use

do not use

Date of creation

Feb 22, 2007 (MN, HD)

Feb 22, 2007 (MN, HD)

Feb 23, 2007 (MN, HD)

Feb 23, 2007 (MN, HD)

Feb 23, 2007 (MN, HD)

Source

CSL EGSE at MPE (Alain)

CSL EGSE at MPE (Alain)

ZEISS EICD

ZEISS EICD

ZEISS EICD

PosLimit

+/- 4.5 deg (13000 units)

+/- 4.5 deg (13000 units)

+/- 5.5 deg (16000 units)

+/- 10.5 deg (28500 units)

+/- 10.5 deg (28500 units)

OutputLimit

+/- 24 mA (6000 units)

+/- 24 mA (6000 units)

+/- 30 mA (7500 units)

+/- 45 mA (11086 units)

+/- 45 mA (11086 units)

Rate

29 units, calculated for chopper pos +/- 12000 units at 100 ms ramp

29 units, calculated for chopper pos +/- 12000 units at 100 ms ramp

29 units, calculated for chopper pos +/- 12000 units at 100 ms ramp

29 units, calculated for chopper pos +/- 12000 units at 100 ms ramp

29 units, calculated for chopper pos +/- 12000 units at 100 ms ramp

Notch filter

480 Hz, DA=2

480 Hz, DA=2

900 Hz, DA=2

900 Hz, DA=2

480 Hz, DA=2

DMC amplification

50.243

50.243

50.09

50.09

50.09

Offset mech. ZP

-1125 units

0 units

0 units

-1195 units

-1195 units

Transition times

not measured

not measured

not measured

not measured

not measured

CSLopt1

CSLopt2-600

CSLopt2-550

CSLopt3-600

CSLopt3-550

Remark

stable fallback parameter set, Warning! Oscillations occur at deflections > 4.1 deg!

Dangerous! Can damage chopper!, worked with CSL EGSE

potentially dangerous, not tested

Working well. This is the default parameter set for now.

potentially dangerous, not tested

Usage

use if other settings fail

never use

do not use

use as default

do not use

Date of creation

Feb 28, 2007 (MN)

Mar 02, 2007 (MN)

Mar 02, 2007 (MN)

Mar 02, 2007 (MN)

Mar 02, 2007 (MN)

Source

CSL EGSE at MPE (Benoit)

CSL EGSE at MPE (Benoit)

CSL EGSE at MPE (Benoit)

CSL EGSE at MPE (Benoit)

CSL EGSE at MPE (Benoit)

PosLimit

+/- 5 deg (14000 units)

+/- 9.5 deg (25000 units)

+/- 9.5 deg (25000 units)

+/- 9.5 deg (25000 units)

+/- 5 deg (14000 units)

OutputLimit

+/- 65 mA (16013 units)

+/- 65 mA (16013 units)

+/- 65 mA (16013 units)

+/- 110 mA (27100 units)

+/- 110 mA (27100 units)

Rate

310 units, calculated for chopper pos +/- 26000 units at 10 ms ramp

310 units, calculated for chopper pos +/- 26000 units at 10 ms ramp

310 units, calculated for chopper pos +/- 26000 units at 10 ms ramp

550 units, calculated for chopper pos +/- 26000 units at 5 ms ramp

550 units, calculated for chopper pos +/- 26000 units at 5 ms ramp

Notch filter

550 Hz, DA=3

600 Hz, DA=3

550 Hz, DA=3

600 Hz, DA=3

550 Hz, DA=3

Offset mech. ZP

-1195 units

-1195 units

-1195 units

-1195 units

-1195 units

Transition time at +/- 4 deg

slow

approx. 25 ms

approx. 25 ms

approx. 17 ms

approx. 17 ms

Transition time at +/- 9 deg

slow

approx. 60 ms

approx. 60 ms

approx. 60 ms

approx. 60 ms


4. Verify function of Ge:Ga detectors, CREs and detector heaters PTD 0.7.3 (TBD!)


  • for detailed test description see PTD_0.7.3 (->file attachment:vergega.pdf) or in CVS

  • prerequisites:
    • PACS in spectroscopic mode
    • blue detector heater should be around 2.1 K
  • short test description:
    • procedures consists of two tests, one in loss-less compression mode to get raw ramps and one in default mode
    • start with test in loss-less compression mode; detector selection table with 14 randomly selected pixel DXS7602 in both channels; execute CUS script SPEC_detector_fmilt073_detsel_llc_OBS.cus (called by tcl-script Generic_CUS_OBS.tcl); script lasts 1659 seconds

    • do then test in default mode; therefore, execute CUS script SPEC_detector_fmilt073_standard_OBS.cus (called by tcl-script Generic_CUS_OBS.tcl); script lasts 134


5. Chopper Synchronization PTD 0.7.6 (TBD!)


5.1. Spectroscopic Mode (TBD!)

  • for detailed test description see PTD_0.7.6_spec (->file attachment:chopsynspec.pdf) or in CVS

  • prerequisites:
    • PACS in spectroscopic mode
    • setup Chopper Configuration Parameter for 8 degree; use from CAL-U table CONFChopper the column FM_sa2_81deg

  • short test description:
    • we use the SPU mode loss less compression
    • detector selection table with 14 randomly selected pixel DXS7602 in both channels

    • use CUS script SPEC_Chopper_fmilt076spec_detsel_llc_OBS.cus for conducting this test (called by tcl-script Generic_CUS_OBS.tcl); script lasts 865 seconds

5.2. Photometeric Mode (TBD!)

  • for detailed test description see PTD_0.7.6_phot (->file attachment:chopsynphot.pdf) or in CVS

  • prerequisites:
    • PACS in photometric mode
    • setup Chopper Configuration Parameter for 8 degree; use from CAL-U table CONFChopper the column FM_sa2_81deg; to do this do following:

      • if chopper controller enabled, disable chopper controller with CUS script XXX (called by tcl-script XXX with parameters XXX)

      • do chopper configuration with CUS script XXX (called by tcl-script XXX with parameters XXX)

      • enable chopper controller with CUS script XXX (called by tcl-scritp XXX with parameters XXX)

  • short test description:
    • we use the SPU mode loss less compression
    • detector selection table with 14 randomly selected pixel DXS7602 in both channels

    • use CUS script PHOT_Chopper_fmilt076phot_detsel_llc_OBS.cus for conducting this test (called by tcl-script Generic_CUS_OBS.tcl); script lasts 51 seconds


6. Internal Calibration Sources PTD 0.7.11_12 & 2.5.3


  • prerequisites:
    • PACS in spectroscopic mode
  • for detailed test description see PTD_0.7.11_12 (->file attachment:cs.pdf) or in CVS

  • the test is divided in several parts:

6.1. Day 16:

  • load up CSs control parameter set using tcl-scripts CONF_CS1_OBS.tcl and CONF_CS2_OBS.tcl and the parameters calUfile=CONFCs1Cs2 and model=FM1

  • heat-up of CSs including switching-on, enabling and setting of temperature: SPEC_CS_fmilt071112_heatup_cs_OBS.cus (Generic_CUS_OBS.tcl)

  • switch-off CSs with SPEC_CS_fmilt071112_switchoff_cs_OBS.cus (Generic_CUS_OBS.tcl)

6.2. Day 17:

  • if necessary change CSs control-parameter in CAL-U table CONFCs1Cs2 -> PTD 2.5.3

  • load up CSs control parameter set using tcl-script CONF_CS1_OBS.tcl and CONF_CS2_OBS.tcl and the parameters calUfile=CONFCs1Cs2 and model=FM1 (TBC)

  • heat-up of CSs including switching-on, enabling and setting of temperature: SPEC_CS_fmilt071112_heatup_cs_OBS.cus (Generic_CUS_OBS.tcl)

  • switch-off CSs with SPEC_CS_fmilt071112_switchoff_cs_OBS.cus (Generic_CUS_OBS.tcl)

6.3. Day 18:

  • this test has two heat-up's and cool-down
    • short test description:
    • switch-on and enable CSs
    • heat-up to CS1 and CS2 temperature value XX/XX (TBD) and stabilization at temperature plateau for in total 80 minutes
    • go to a second temperature plateau (CS1=XX/CS2=XX;TBD) - 60 minutes
    • cool-down (CS1=XX/CS2=XX;TBD) - script stops then 60 seconds after commanding the cool-down values; stop of recording file should be done after 60(TBD) minutes
  • Conduct following steps at commanding concole:
    • load up CSs control parameter set using tcl-scripts CONF_CS1_OBS.tcl and CONF_CS2_OBS.tcl and the parameters calUfile=CONFCs1Cs2 and model=FM1 (TBC)

    • use CUS script SPEC_CS_fmilt071112_swon_cs_perf_OBS.cus for conducting this test (called by tcl-script) Generic_CUS_OBS.tcl)

    • CSs will not be switched-off
  • comment:
    • if necessary (e.g. in phase of control loop optimization) the temperature values in the CUS-script should be adapted


7. Optimum Bias Setting PTD 1.2.1_2_17


  • set OGSE BB Temperatures (BB1=25.3K;BB2=24.7K) with OGSEBBSetUp_OBS.tcl
  • verify flip mirrors if on correct position
  • wait for stabilization
  • set OGSE Chopper Wheel (CW;ROT;438mHz) with OGSEOpticsSetUp_OBS.tcl
  • set spectroscopic housekeeping (e.g., use PACS_GeGa_SFT_Init_Cold_He2_OBS - generic)

  • upload proper control parameter's using tcl-script CONF_chopper_OBS.tcl with parameters fileU=CONFChopper and calUcol=ZEISSoffset2

  • switch-on chopper and check if it is stable
  • enable chopper and check if it it is stable
  • move chopper to Pos 0 manually
  • switch-on and enable grating
  • heater to 1.5mA
  • use SPEC_detector_fmilt121_part1_OBS for red array (1648 sec)

  • after end of test which lasts about 1648 sec perhaps switch-off chopper wheel
  • put FW to Pos B

Instruction for Thomas:

  • put FW to Pos B
  • set OGSE BBs to new temp (BB1=32.75K;BB2=31.25K) with OGSEBBSetUp_OBS.tcl
  • wait for stabilization
  • set OGSE Chopper Wheel (CW;ROT;438mHz) with OGSEOpticsSetUp_OBS.tcl
  • set spectroscopic housekeeping (e.g., use PACS_GeGa_SFT_Init_Cold_He2_OBS - generic)

  • upload proper control parameter's using tcl-script CONF_chopper_OBS.tcl with parameters fileU=CONFChopper and calUcol=ZEISSoffset2

  • switch-on chopper and check if it is stable
  • enable chopper and check if it it is stable
  • move chopper to Pos 0 manually
  • switch-on and enable grating
  • use SPEC_detector_fmilt121_part2_OBS for blue array (12198 sec)

  • after end of the script put OGSE BBs down to 4K with OGSEBBSetUp_OBS.tcl and switch-off Chopper wheel with OGSEOpticsSetUp_OBS.tcl BV: check grating position

    HD&BV: get from PR and RH values for detector heater and stabilization time

    • Heater temperature stabilisation time: 15min
      • CUS script changed: 10 minutes wait + 8 minutes Before analysis, check that temperature and signal are constant over the 8 min.
    • Blue Heater conversions Kelvin - mA :

      Heater temperature K

      Heater current mA

      2.1

      0.54

      2.3

      1.10

      2.5

      1.51

      2.7

      1.84

      2.9

      2.15


8. CRE Check Out Voltage PTD 1.2.4


  • set spectroscopic housekeeping (e.g., use PACS_GeGa_SFT_Init_Cold_He2_OBS - generic)

  • upload proper control parameter's using tcl-script CONF_chopper_OBS.tcl with parameters fileU=CONFChopper and calUcol=CSLopt1

  • start diag. HK using SPEC_Chopper_dhk_4hk_1khz_OBS.cus (Generic_CUS_script.tcl)

  • switch-on chopper and check if it is stable
  • enable chopper and check if it it is stable
  • stop diag. HK using SPEC_Chopper_dhk_stop_OBS.cus (Generic_CUS_script.tcl)

  • start SPEC_detector_fmilt124_30mV_detsel_llc_OBS.cus (Generic)

  • start SPEC_detector_fmilt124_60mV_detsel_llc_OBS.cus (Generic)

  • if necessary disable and switch-off chopper

Herschel: PACS/FM_ILT_Procedures/MPIAprocedures (last edited 2009-07-15 14:32:36 by localhost)