Differences between revisions 50 and 51
Revision 50 as of 2007-02-07 17:51:10
Size: 4898
Comment:
Revision 51 as of 2007-02-24 08:56:19
Size: 5439
Comment:
Deletions are marked like this. Additions are marked like this.
Line 88: Line 88:
------------------------------------------------------------------------------------
= Internal Calibration Sources PTD 0.7.11_12 & 2.5.3 (TBD!) =
------------------------------------------------------------------------------------
== Spectroscopic Mode ==

 * for detailed test description see '''PTD_0.7.11_12''' (->file attachment:cs.pdf) or in CVS

 * prerequisites:
   * PACS in spectroscopic mode
   * setup Chopper Configuration Parameter for 8 degree; use from CAL-U table '''CONFChopper''' the column '''FM_sa2_81deg'''

H. Dannerbauer; 07-Dec-2006

Index TableOfContents(2)


1. Determination of Chopper Offset


  • for detailed test description see PTD_2.3.1 (->file attachment:chopper.pdf) or in CVS

  • Chopper has to be switched-off and controller disabled
  • start PacketRecorder

  • start diag. HK with SPEC_Chopper_dhk_imt_OBS using Generic_CUS_script.tcl

  • determine chopper offset (chop_offset) from DM_CHOP_CUR_POS
  • include the new chopper offset in chopper control parameter set
  • stop diag. HK with SPEC_Chopper_dhk_stop_OBS using Generic_CUS_script.tcl

  • close PacketRecorder

  • switch-on chopper
  • load up chopper control parameters including the new offset
    • -> tcl-script CONF_chopper_OBS.tcl with parameters fileU and calUcol

  • enable chopper controller in order to upload new control parameter set
    • -> EnDis_chopper_OBS.tcl and parameter onoff ON


2. Chopper Open Loop Measurements


  • for detailed test description see PTD_2.3.1 (->file attachment:chopper.pdf) or in CVS

  • chopper controller has to be enabled
  • start PacketRecorder

  • start script SPEC_Chopper_OpenLoop_MoveAbs_OBS.tcl given following parameters: chop_max_ma (maximum current in mA), chop_step_ma (current step width in mA) and chop_damptime (plateau duration in sec)

  • close PacketRecorder

  • if neccessary disable and switch-off chopper


3. Verify function of Ge:Ga detectors, CREs and detector heaters PTD 0.7.3 (TBD!)


  • for detailed test description see PTD_0.7.3 (->file attachment:vergega.pdf) or in CVS

  • prerequisites:
    • PACS in spectroscopic mode
    • blue detector heater should be around 2.1 K
  • short test description:
    • procedures consists of two tests, one in loss-less compression mode to get raw ramps and one in default mode
    • start with test in loss-less compression mode; detector selection table with 14 randomly selected pixel DXS7602 in both channels; execute CUS script SPEC_detector_fmilt073_detsel_llc_OBS.cus (called by tcl-script Generic_CUS_OBS.tcl); script lasts 1659 seconds

    • do then test in default mode; therefore, execute CUS script SPEC_detector_fmilt073_standard_OBS.cus (called by tcl-script Generic_CUS_OBS.tcl); script lasts 134


4. Chopper Synchronization PTD 0.7.6 (TBD!)


4.1. Spectroscopic Mode (TBD!)

  • for detailed test description see PTD_0.7.6_spec (->file attachment:chopsynspec.pdf) or in CVS

  • prerequisites:
    • PACS in spectroscopic mode
    • setup Chopper Configuration Parameter for 8 degree; use from CAL-U table CONFChopper the column FM_sa2_81deg

  • short test description:
    • we use the SPU mode loss less compression
    • detector selection table with 14 randomly selected pixel DXS7602 in both channels

    • use CUS script SPEC_Chopper_fmilt076spec_detsel_llc_OBS.cus for conducting this test (called by tcl-script Generic_CUS_OBS.tcl); script lasts 865 seconds

4.2. Photometeric Mode (TBD!)

  • for detailed test description see PTD_0.7.6_phot (->file attachment:chopsynphot.pdf) or in CVS

  • prerequisites:
    • PACS in photometric mode
    • setup Chopper Configuration Parameter for 8 degree; use from CAL-U table CONFChopper the column FM_sa2_81deg; to do this do following:

      • if chopper controller enabled, disable chopper controller with CUS script XXX (called by tcl-script XXX with parameters XXX)

      • do chopper configuration with CUS script XXX (called by tcl-script XXX with parameters XXX)

      • enable chopper controller with CUS script XXX (called by tcl-scritp XXX with parameters XXX)

  • short test description:
    • we use the SPU mode loss less compression
    • detector selection table with 14 randomly selected pixel DXS7602 in both channels

    • use CUS script PHOT_Chopper_fmilt076phot_detsel_llc_OBS.cus for conducting this test (called by tcl-script Generic_CUS_OBS.tcl); script lasts 51 seconds


5. Internal Calibration Sources PTD 0.7.11_12 & 2.5.3 (TBD!)


5.1. Spectroscopic Mode

  • for detailed test description see PTD_0.7.11_12 (->file attachment:cs.pdf) or in CVS

  • prerequisites:
    • PACS in spectroscopic mode
    • setup Chopper Configuration Parameter for 8 degree; use from CAL-U table CONFChopper the column FM_sa2_81deg

Herschel: PACS/FM_ILT_Procedures/MPIAprocedures (last edited 2009-07-15 14:32:36 by localhost)