Size: 77
Comment:
|
Size: 3940
Comment:
|
Deletions are marked like this. | Additions are marked like this. |
Line 1: | Line 1: |
= Chopper Synchronization = | #pragma section-numbers on |
Line 3: | Line 3: |
= Photometer Mode = | H. Dannerbauer; 07-Dec-2006 |
Line 5: | Line 6: |
= Spectrometer Mode = | '''Index''' [[TableOfContents(2)]] ------------------------------------------------------------------------------------ = Determination of Chopper Offset = ------------------------------------------------------------------------------------ * Chopper has to be switched-off and controller disabled * start PacketRecorder * start diag. HK with '''SPEC_Chopper_dhk_imt_OBS''' * from DM_CHOP_CUR_POS the chopper offset (chop_offfset) will be determined * include the new chopper offset in chopper control parameter set -> PosOffset=(-*1)*chop_offset * stop diag. HK witth '''SPEC_Chopper_dhk_stop_OBS''' * close PacketRecorder * switch-on Chopper * load up Chopper control parameters including the new offset -> tcl-script '''CONF_chopper_OBS.tcl''' with parameter name of CAL-U and control parameter set ------------------------------------------------------------------------------------ = Verify function of Ge:Ga detectors, CREs and detector heaters PTD 0.7.3 (TBD!) = ------------------------------------------------------------------------------------ * for detailed test description see '''PTD_0.7.3''' (->file attachment:vergega.pdf) or in CVS * prerequisites: * PACS in spectroscopic mode * blue detector heater should be around 2.1 K * short test description: * procedures consists of two tests, one in loss-less compression mode to get raw ramps and one in default mode * start with test in loss-less compression mode; detector selection table with 14 randomly selected pixel '''DXS7602''' in both channels; execute CUS script '''SPEC_detector_fmilt073_detsel_llc_OBS.cus''' (called by tcl-script '''Generic_CUS_OBS.tcl'''); script lasts 1659 seconds * do then test in default mode; therefore, execute CUS script '''SPEC_detector_fmilt073_standard_OBS.cus''' (called by tcl-script '''Generic_CUS_OBS.tcl'''); script lasts 134 ------------------------------------------------------------------------------------ = Chopper Synchronization PTD 0.7.6 (TBD!) = ------------------------------------------------------------------------------------ == Spectroscopic Mode (TBD!) == * for detailed test description see '''PTD_0.7.6_spec''' (->file attachment:chopsynspec.pdf) or in CVS * prerequisites: * PACS in spectroscopic mode * setup Chopper Configuration Parameter for 8 degree; use from CAL-U table '''CONFChopper''' the column '''FM_sa2_81deg''' * short test description: * we use the SPU mode loss less compression * detector selection table with 14 randomly selected pixel '''DXS7602''' in both channels * use CUS script '''SPEC_Chopper_fmilt076spec_detsel_llc_OBS.cus''' for conducting this test (called by tcl-script '''Generic_CUS_OBS.tcl'''); script lasts 865 seconds == Photometeric Mode (TBD!) == * for detailed test description see '''PTD_0.7.6_phot''' (->file attachment:chopsynphot.pdf) or in CVS * prerequisites: * PACS in photometric mode * setup Chopper Configuration Parameter for 8 degree; use from CAL-U table '''CONFChopper''' the column '''FM_sa2_81deg'''; to do this do following: * if chopper controller enabled, disable chopper controller with CUS script '''XXX''' (called by tcl-script '''XXX''' with parameters XXX) * do chopper configuration with CUS script '''XXX''' (called by tcl-script '''XXX''' with parameters '''XXX''') * enable chopper controller with CUS script '''XXX''' (called by tcl-scritp '''XXX''' with parameters '''XXX''') * short test description: * we use the SPU mode loss less compression * detector selection table with 14 randomly selected pixel '''DXS7602''' in both channels * use CUS script '''PHOT_Chopper_fmilt076phot_detsel_llc_OBS.cus''' for conducting this test (called by tcl-script '''Generic_CUS_OBS.tcl'''); script lasts 51 seconds |
H. Dannerbauer; 07-Dec-2006
Index TableOfContents(2)
1. Determination of Chopper Offset
- Chopper has to be switched-off and controller disabled
start PacketRecorder
start diag. HK with SPEC_Chopper_dhk_imt_OBS
- from DM_CHOP_CUR_POS the chopper offset (chop_offfset) will be determined
- include the new chopper offset in chopper control parameter set
-> PosOffset=(-*1)*chop_offset
stop diag. HK witth SPEC_Chopper_dhk_stop_OBS
close PacketRecorder
- switch-on Chopper
- load up Chopper control parameters including the new offset
-> tcl-script CONF_chopper_OBS.tcl with parameter name of CAL-U and control parameter set
2. Verify function of Ge:Ga detectors, CREs and detector heaters PTD 0.7.3 (TBD!)
for detailed test description see PTD_0.7.3 (->file attachment:vergega.pdf) or in CVS
- prerequisites:
- PACS in spectroscopic mode
- blue detector heater should be around 2.1 K
- short test description:
- procedures consists of two tests, one in loss-less compression mode to get raw ramps and one in default mode
start with test in loss-less compression mode; detector selection table with 14 randomly selected pixel DXS7602 in both channels; execute CUS script SPEC_detector_fmilt073_detsel_llc_OBS.cus (called by tcl-script Generic_CUS_OBS.tcl); script lasts 1659 seconds
do then test in default mode; therefore, execute CUS script SPEC_detector_fmilt073_standard_OBS.cus (called by tcl-script Generic_CUS_OBS.tcl); script lasts 134
3. Chopper Synchronization PTD 0.7.6 (TBD!)
3.1. Spectroscopic Mode (TBD!)
for detailed test description see PTD_0.7.6_spec (->file attachment:chopsynspec.pdf) or in CVS
- prerequisites:
- PACS in spectroscopic mode
setup Chopper Configuration Parameter for 8 degree; use from CAL-U table CONFChopper the column FM_sa2_81deg
- short test description:
- we use the SPU mode loss less compression
detector selection table with 14 randomly selected pixel DXS7602 in both channels
use CUS script SPEC_Chopper_fmilt076spec_detsel_llc_OBS.cus for conducting this test (called by tcl-script Generic_CUS_OBS.tcl); script lasts 865 seconds
3.2. Photometeric Mode (TBD!)
for detailed test description see PTD_0.7.6_phot (->file attachment:chopsynphot.pdf) or in CVS
- prerequisites:
- PACS in photometric mode
setup Chopper Configuration Parameter for 8 degree; use from CAL-U table CONFChopper the column FM_sa2_81deg; to do this do following:
if chopper controller enabled, disable chopper controller with CUS script XXX (called by tcl-script XXX with parameters XXX)
do chopper configuration with CUS script XXX (called by tcl-script XXX with parameters XXX)
enable chopper controller with CUS script XXX (called by tcl-scritp XXX with parameters XXX)
- short test description:
- we use the SPU mode loss less compression
detector selection table with 14 randomly selected pixel DXS7602 in both channels
use CUS script PHOT_Chopper_fmilt076phot_detsel_llc_OBS.cus for conducting this test (called by tcl-script Generic_CUS_OBS.tcl); script lasts 51 seconds