Differences between revisions 19 and 20
Revision 19 as of 2006-12-06 14:45:33
Size: 1526
Comment:
Revision 20 as of 2006-12-06 14:47:27
Size: 1530
Comment:
Deletions are marked like this. Additions are marked like this.
Line 28: Line 28:
   attachment:chopsynphot    attachment:chopsynphot.pdf

H. Dannerbauer; 06-Dec-2006

Index TableOfContents(2)


1. Chopper Synchronization PTD 0.7.6


1.1. Spectroscopic Mode

  • for detailed test description see PTD_0.7.6_spec in CVS

  • prerequisites
    • PACS in spectroscopic mode
    • setup Chopper Configuration Parameter for 8 degree; use from CAL-U table CONFChopper the column FM_sa1_81deg

  • short test description
    • we use the SPU mode loss less compression
    • detector selection table with 14 randomly selected pixel DXS7602 in both channels

    • use CUS script SPEC_Chopper_fmilt076spec_detsel_llc_OBS.cus for conducting this test (called by tcl-script Generic_CUS_OBS.tcl)

1.2. Photometeric Mode

  • for detailed test description see PTD_0.7.6_phot in CVS

    • attachment:chopsynphot.pdf
  • prerequisites
    • PACS in photometric mode
    • setup Chopper Configuration Parameter for 8 degree; use from CAL-U table CONFChopper the column FM_sa1_81deg

  • short test description
    • we use the SPU mode loss less compression
    • detector selection table with 14 randomly selected pixel DXS7602 in both channels

    • use CUS script PHOT_Chopper_fmilt076phot_detsel_llc_OBS.cus for conducting this test (called by tcl-script Generic_CUS_OBS.tcl)

Herschel: PACS/FM_ILT_Procedures/MPIAprocedures (last edited 2009-07-15 14:32:36 by localhost)