Line 1: Line 1:
Test #pragma section-numbers on

H. Dannerbauer;
07-Dec-2006

'''Index'''
[[TableOfContents(2)]]
------------------------------------------------------------------------------------
= Determination of Chopper Offset =
------------------------------------------------------------------------------------

 * for detailed test description see '''PTD_2.3.1''' (->file attachment:chopper.pdf) or in CVS

 * Chopper has to be switched-off and controller disabled
 * upload Chopper CPRs with position offset of zero using tcl-script '''CONF_Chopper_OBS.tcl''' with parameters '''fileU''' and '''calUcol'''
 * start PacketRecorder
 * start diag. HK with '''SPEC_Chopper_dhk_imt_OBS''' using '''Generic_CUS_script.tcl'''
 * determine chopper offset (chop_offset) from DM_CHOP_CUR_POS
 * include the new chopper offset in chopper control parameter set
   -> PosOffset=(-1)*chop_offset
 * stop diag. HK with '''SPEC_Chopper_dhk_stop_OBS''' using '''Generic_CUS_script.tcl'''
 * close PacketRecorder
 * switch-on chopper
 * load up chopper control parameters including the new offset
   -> tcl-script '''CONF_chopper_OBS.tcl''' with parameters '''fileU''' and '''calUcol'''
 * enable chopper controller in order to upload new control parameter set
  -> '''EnDis_chopper_OBS.tcl''' and parameter '''onoff''' ON

------------------------------------------------------------------------------------
= Chopper Open Loop Measurements =
------------------------------------------------------------------------------------
 
 * for detailed test description see '''PTD_2.3.1''' (->file attachment:chopper.pdf) or in CVS

 * it is recommended to upload proper control parameter's using tcl-script '''CONF_chopper_OBS.tcl''' with parameters '''fileU''' and '''calUcol'''
 * test should be operated in nominal spectroscopic mode; to do this execute the GeGa SFT; e.g. for He2 use '''PACS_GeGa_SFT_Init_Cold_He2_OBS.cus''' (Generic_CUS_script.tcl)
 * chopper controller has to be at least disabled
 * start diag. HK using '''SPEC_Chopper_dhk_4hk_1khz_OBS.cus''' (Generic_CUS_script.tcl)
 * start PacketRecorder
 * start script '''SPEC_Chopper_OpenLoop_MoveAbs_OBS.tcl''' given following parameters: '''chop_max_ma''' (maximum current in mA), '''chop_step_ma''' (current step width in mA) and '''chop_damptime''' (plateau duration in sec)
 * close PacketRecorder
 * stop diag. HK using '''SPEC_Chopper_dhk_stop_OBS.cus''' (Generic_CUS_script.tcl)
 * at the end of the test chopper controller is disabled; if necessary switch-off chopper

------------------------------------------------------------------------------------
= Verify function of Ge:Ga detectors, CREs and detector heaters PTD 0.7.3 (TBD!) =
------------------------------------------------------------------------------------

 * for detailed test description see '''PTD_0.7.3''' (->file attachment:vergega.pdf) or in CVS

 * prerequisites:
   * PACS in spectroscopic mode
   * blue detector heater should be around 2.1 K
   
 * short test description:
   * procedures consists of two tests, one in loss-less compression mode to get raw ramps and one in default mode
   * start with test in loss-less compression mode; detector selection table with 14 randomly selected pixel '''DXS7602''' in both channels; execute CUS script '''SPEC_detector_fmilt073_detsel_llc_OBS.cus''' (called by tcl-script '''Generic_CUS_OBS.tcl'''); script lasts 1659 seconds
   * do then test in default mode; therefore, execute CUS script '''SPEC_detector_fmilt073_standard_OBS.cus''' (called by tcl-script '''Generic_CUS_OBS.tcl'''); script lasts 134

------------------------------------------------------------------------------------
= Chopper Synchronization PTD 0.7.6 (TBD!) =
------------------------------------------------------------------------------------
== Spectroscopic Mode (TBD!) ==

 * for detailed test description see '''PTD_0.7.6_spec''' (->file attachment:chopsynspec.pdf) or in CVS

 * prerequisites:
   * PACS in spectroscopic mode
   * setup Chopper Configuration Parameter for 8 degree; use from CAL-U table '''CONFChopper''' the column '''FM_sa2_81deg'''

 * short test description:
   * we use the SPU mode loss less compression
   * detector selection table with 14 randomly selected pixel '''DXS7602''' in both channels
   * use CUS script '''SPEC_Chopper_fmilt076spec_detsel_llc_OBS.cus''' for conducting this test (called by tcl-script '''Generic_CUS_OBS.tcl'''); script lasts 865 seconds


== Photometeric Mode (TBD!) ==

 * for detailed test description see '''PTD_0.7.6_phot''' (->file attachment:chopsynphot.pdf) or in CVS

 * prerequisites:
   * PACS in photometric mode
   * setup Chopper Configuration Parameter for 8 degree; use from CAL-U table '''CONFChopper''' the column '''FM_sa2_81deg'''; to do this do following:
     * if chopper controller enabled, disable chopper controller with CUS script '''XXX''' (called by tcl-script '''XXX''' with parameters XXX)
     * do chopper configuration with CUS script '''XXX''' (called by tcl-script '''XXX''' with parameters '''XXX''')
     * enable chopper controller with CUS script '''XXX''' (called by tcl-scritp '''XXX''' with parameters '''XXX''')

 * short test description:
   * we use the SPU mode loss less compression
   * detector selection table with 14 randomly selected pixel '''DXS7602''' in both channels
   * use CUS script '''PHOT_Chopper_fmilt076phot_detsel_llc_OBS.cus''' for conducting this test (called by tcl-script '''Generic_CUS_OBS.tcl'''); script lasts 51 seconds
 
------------------------------------------------------------------------------------
= Internal Calibration Sources PTD 0.7.11_12 & 2.5.3 =
------------------------------------------------------------------------------------
 * prerequisites:
   * PACS in spectroscopic mode

 * for detailed test description see '''PTD_0.7.11_12''' (->file attachment:cs.pdf) or in CVS

 * the test is divided in several parts:

== Day 16: ==
 * load up CSs control parameter set using tcl-scripts '''CONF_CS1_OBS.tcl''' and '''CONF_CS2_OBS.tcl''' and the parameters '''calUfile=CONFCs1Cs2''' and '''model=FM1'''
 * heat-up of CSs including switching-on, enabling and setting of temperature: '''SPEC_CS_fmilt071112_heatup_cs_OBS.cus''' (Generic_CUS_OBS.tcl)
 * switch-off CSs with '''SPEC_CS_fmilt071112_switchoff_cs_OBS.cus''' (Generic_CUS_OBS.tcl)

== Day 17: ==
 * if necessary change CSs control-parameter in CAL-U table '''CONFCs1Cs2''' -> PTD 2.5.3
 * load up CSs control parameter set using tcl-script '''CONF_CS1_OBS.tcl''' and '''CONF_CS2_OBS.tcl''' and the parameters '''calUfile=CONFCs1Cs2''' and '''model=FM1''' (TBC)
 * heat-up of CSs including switching-on, enabling and setting of temperature: '''SPEC_CS_fmilt071112_heatup_cs_OBS.cus''' (Generic_CUS_OBS.tcl)
 * switch-off CSs with '''SPEC_CS_fmilt071112_switchoff_cs_OBS.cus''' (Generic_CUS_OBS.tcl)

== Day 18: ==
 * this test has two heat-up's and cool-down
   * short test description:
   * switch-on and enable CSs
   * heat-up to CS1 and CS2 temperature value XX/XX (TBD) and stabilization at temperature plateau for in total 80 minutes
   * go to a second temperature plateau (CS1=XX/CS2=XX;TBD) - 60 minutes
   * cool-down (CS1=XX/CS2=XX;TBD) - script stops then 60 seconds after commanding the cool-down values; stop of recording file should be done after 60(TBD) minutes

 * Conduct following steps at commanding concole:
   * load up CSs control parameter set using tcl-scripts '''CONF_CS1_OBS.tcl''' and '''CONF_CS2_OBS.tcl''' and the parameters '''calUfile=CONFCs1Cs2''' and '''model=FM1''' (TBC)
   * use CUS script '''SPEC_CS_fmilt071112_swon_cs_perf_OBS.cus''' for conducting this test (called by tcl-script) '''Generic_CUS_OBS.tcl''')
   * CSs will not be switched-off

 * comment:
   if necessary (e.g. in phase of control loop optimization) the temperature values in the CUS-script should be adapted

------------------------------------------------------------------------------------
= Optimum Bias Setting PTD 1.2.1_2_17 =
------------------------------------------------------------------------------------

 * set OGSE BB Temperatures (BB1=25.3K;BB2=24.7K)
 * wait for stabilization
 * set OGSE Chopper Wheel (CW;ROT;438mHz)
 * set spectroscopic housekeeping (e.g., use PACS_GeGa_SFT...He2)
 * upload proper control parameter's using tcl-script '''CONF_chopper_OBS.tcl''' with parameters '''fileU=CONFChopper''' and '''calUcol=ZEISSoffset2'''
 * switch-on chopper and check if it is stable
 * enable chopper and check if it it is stable
 * switch-on and enable grating
 * heater to 1.5mA
 * use '''SPEC_detector_fmilt121_part1_OBS''' for red array (1648 sec)
 * after end of test which lasts about 1648 sec perhaps switch-off chopper wheel
 * put FW to Pos B
 * set OGSE BBs to new temp (BB1=32.45K;BB2=31.55K)
 * use '''SPEC_detector_fmilt121_part2_OBS''' for blue array (8035 sec)
 * switch-off OGSE BBs and Chopper wheel

H. Dannerbauer; 07-Dec-2006

Index TableOfContents(2)


1. Determination of Chopper Offset


  • for detailed test description see PTD_2.3.1 (->file attachment:chopper.pdf) or in CVS

  • Chopper has to be switched-off and controller disabled
  • upload Chopper CPRs with position offset of zero using tcl-script CONF_Chopper_OBS.tcl with parameters fileU and calUcol

  • start PacketRecorder

  • start diag. HK with SPEC_Chopper_dhk_imt_OBS using Generic_CUS_script.tcl

  • determine chopper offset (chop_offset) from DM_CHOP_CUR_POS
  • include the new chopper offset in chopper control parameter set
  • stop diag. HK with SPEC_Chopper_dhk_stop_OBS using Generic_CUS_script.tcl

  • close PacketRecorder

  • switch-on chopper
  • load up chopper control parameters including the new offset
    • -> tcl-script CONF_chopper_OBS.tcl with parameters fileU and calUcol

  • enable chopper controller in order to upload new control parameter set
    • -> EnDis_chopper_OBS.tcl and parameter onoff ON


2. Chopper Open Loop Measurements


  • for detailed test description see PTD_2.3.1 (->file attachment:chopper.pdf) or in CVS

  • it is recommended to upload proper control parameter's using tcl-script CONF_chopper_OBS.tcl with parameters fileU and calUcol

  • test should be operated in nominal spectroscopic mode; to do this execute the GeGa SFT; e.g. for He2 use PACS_GeGa_SFT_Init_Cold_He2_OBS.cus (Generic_CUS_script.tcl)

  • chopper controller has to be at least disabled
  • start diag. HK using SPEC_Chopper_dhk_4hk_1khz_OBS.cus (Generic_CUS_script.tcl)

  • start PacketRecorder

  • start script SPEC_Chopper_OpenLoop_MoveAbs_OBS.tcl given following parameters: chop_max_ma (maximum current in mA), chop_step_ma (current step width in mA) and chop_damptime (plateau duration in sec)

  • close PacketRecorder

  • stop diag. HK using SPEC_Chopper_dhk_stop_OBS.cus (Generic_CUS_script.tcl)

  • at the end of the test chopper controller is disabled; if necessary switch-off chopper


3. Verify function of Ge:Ga detectors, CREs and detector heaters PTD 0.7.3 (TBD!)


  • for detailed test description see PTD_0.7.3 (->file attachment:vergega.pdf) or in CVS

  • prerequisites:
    • PACS in spectroscopic mode
    • blue detector heater should be around 2.1 K
  • short test description:
    • procedures consists of two tests, one in loss-less compression mode to get raw ramps and one in default mode
    • start with test in loss-less compression mode; detector selection table with 14 randomly selected pixel DXS7602 in both channels; execute CUS script SPEC_detector_fmilt073_detsel_llc_OBS.cus (called by tcl-script Generic_CUS_OBS.tcl); script lasts 1659 seconds

    • do then test in default mode; therefore, execute CUS script SPEC_detector_fmilt073_standard_OBS.cus (called by tcl-script Generic_CUS_OBS.tcl); script lasts 134


4. Chopper Synchronization PTD 0.7.6 (TBD!)


4.1. Spectroscopic Mode (TBD!)

  • for detailed test description see PTD_0.7.6_spec (->file attachment:chopsynspec.pdf) or in CVS

  • prerequisites:
    • PACS in spectroscopic mode
    • setup Chopper Configuration Parameter for 8 degree; use from CAL-U table CONFChopper the column FM_sa2_81deg

  • short test description:
    • we use the SPU mode loss less compression
    • detector selection table with 14 randomly selected pixel DXS7602 in both channels

    • use CUS script SPEC_Chopper_fmilt076spec_detsel_llc_OBS.cus for conducting this test (called by tcl-script Generic_CUS_OBS.tcl); script lasts 865 seconds

4.2. Photometeric Mode (TBD!)

  • for detailed test description see PTD_0.7.6_phot (->file attachment:chopsynphot.pdf) or in CVS

  • prerequisites:
    • PACS in photometric mode
    • setup Chopper Configuration Parameter for 8 degree; use from CAL-U table CONFChopper the column FM_sa2_81deg; to do this do following:

      • if chopper controller enabled, disable chopper controller with CUS script XXX (called by tcl-script XXX with parameters XXX)

      • do chopper configuration with CUS script XXX (called by tcl-script XXX with parameters XXX)

      • enable chopper controller with CUS script XXX (called by tcl-scritp XXX with parameters XXX)

  • short test description:
    • we use the SPU mode loss less compression
    • detector selection table with 14 randomly selected pixel DXS7602 in both channels

    • use CUS script PHOT_Chopper_fmilt076phot_detsel_llc_OBS.cus for conducting this test (called by tcl-script Generic_CUS_OBS.tcl); script lasts 51 seconds


5. Internal Calibration Sources PTD 0.7.11_12 & 2.5.3


  • prerequisites:
    • PACS in spectroscopic mode
  • for detailed test description see PTD_0.7.11_12 (->file attachment:cs.pdf) or in CVS

  • the test is divided in several parts:

5.1. Day 16:

  • load up CSs control parameter set using tcl-scripts CONF_CS1_OBS.tcl and CONF_CS2_OBS.tcl and the parameters calUfile=CONFCs1Cs2 and model=FM1

  • heat-up of CSs including switching-on, enabling and setting of temperature: SPEC_CS_fmilt071112_heatup_cs_OBS.cus (Generic_CUS_OBS.tcl)

  • switch-off CSs with SPEC_CS_fmilt071112_switchoff_cs_OBS.cus (Generic_CUS_OBS.tcl)

5.2. Day 17:

  • if necessary change CSs control-parameter in CAL-U table CONFCs1Cs2 -> PTD 2.5.3

  • load up CSs control parameter set using tcl-script CONF_CS1_OBS.tcl and CONF_CS2_OBS.tcl and the parameters calUfile=CONFCs1Cs2 and model=FM1 (TBC)

  • heat-up of CSs including switching-on, enabling and setting of temperature: SPEC_CS_fmilt071112_heatup_cs_OBS.cus (Generic_CUS_OBS.tcl)

  • switch-off CSs with SPEC_CS_fmilt071112_switchoff_cs_OBS.cus (Generic_CUS_OBS.tcl)

5.3. Day 18:

  • this test has two heat-up's and cool-down
    • short test description:
    • switch-on and enable CSs
    • heat-up to CS1 and CS2 temperature value XX/XX (TBD) and stabilization at temperature plateau for in total 80 minutes
    • go to a second temperature plateau (CS1=XX/CS2=XX;TBD) - 60 minutes
    • cool-down (CS1=XX/CS2=XX;TBD) - script stops then 60 seconds after commanding the cool-down values; stop of recording file should be done after 60(TBD) minutes
  • Conduct following steps at commanding concole:
    • load up CSs control parameter set using tcl-scripts CONF_CS1_OBS.tcl and CONF_CS2_OBS.tcl and the parameters calUfile=CONFCs1Cs2 and model=FM1 (TBC)

    • use CUS script SPEC_CS_fmilt071112_swon_cs_perf_OBS.cus for conducting this test (called by tcl-script) Generic_CUS_OBS.tcl)

    • CSs will not be switched-off
  • comment:
    • if necessary (e.g. in phase of control loop optimization) the temperature values in the CUS-script should be adapted


6. Optimum Bias Setting PTD 1.2.1_2_17


  • set OGSE BB Temperatures (BB1=25.3K;BB2=24.7K)
  • wait for stabilization
  • set OGSE Chopper Wheel (CW;ROT;438mHz)
  • set spectroscopic housekeeping (e.g., use PACS_GeGa_SFT...He2)

  • upload proper control parameter's using tcl-script CONF_chopper_OBS.tcl with parameters fileU=CONFChopper and calUcol=ZEISSoffset2

  • switch-on chopper and check if it is stable
  • enable chopper and check if it it is stable
  • switch-on and enable grating
  • heater to 1.5mA
  • use SPEC_detector_fmilt121_part1_OBS for red array (1648 sec)

  • after end of test which lasts about 1648 sec perhaps switch-off chopper wheel
  • put FW to Pos B
  • set OGSE BBs to new temp (BB1=32.45K;BB2=31.55K)
  • use SPEC_detector_fmilt121_part2_OBS for blue array (8035 sec)

  • switch-off OGSE BBs and Chopper wheel

Herschel: PACS/FM_ILT_Procedures/MPIAprocedures (last edited 2009-07-15 14:32:36 by localhost)