Differences between revisions 7 and 8
Revision 7 as of 2007-03-03 10:06:16
Size: 2586
Comment:
Revision 8 as of 2007-03-03 10:06:39
Size: 2580
Comment:
Deletions are marked like this. Additions are marked like this.
Line 30: Line 30:
     || - 8.5 deg || || -22730 || center CS1 ||             || - 8.5 deg || || -22730 || center CS1 ||

FM ILT reference numbers

Ge:Ga detector settings

  • Blue
    • par

      value

      comment

      Bias d

      198 mV

      From bias / heater loop test

      Bias r

      10 mV

      Component level value - to be verified on CRE checkout test

      Heater

      1.5 mA

      Result from bias / heater loop test (2.68K/1.84mA) exceeds thermal budget

  • Red
    • par

      value

      comment

      Bias d

      69 mV

      From bias / heater loop test

      Bias r

      10 mV

      Component level value - to be verified on CRE checkout test

Grating

  • Configuration table: CONFGrat. Column : FMILTFast / FMILT_R_Fast for redundant
  • Accessible range :
    • 45K - 1045K (nominal)
    • 65K - 1065K (redundant)
  • Conversion wavelength - grating position:

Chopper

  • !!! Do NOT use the chopper in redundant mode !!!
  • Configuration table CONFChopper / column : CSLopt3-600
  • Accessible range : -23000 to +23000
    • At +/- 25000 a field plate limit disables the controller
  • Conversion chopper position - angle on the sky - FPU angle :
    • FPU Angle

      Sky Angle

      Chopper pos

      Comment

      ||

      - 8.5 deg || || -22730 || center CS1 ||

      -4 deg

      -11451

      0 deg

      0

      650

      Optical zero

      +4 deg

      12431

      + 8.5 deg

      23000

      center CS2

  • Other angles - in the PCSS (release >20070303A) :

    •           from all import *
                # FPU angle : offset 650
                print 650 + fpuAngle2ChopPosPol(Double1d([-4, 0, 4]))
                # Sky angle : scaling factor 80.69
                print (650 + fpuAngle2ChopPosPol(Double1d([-4, 0, 4])))/80.69

Filter Wheel PHOT

  • POS A = Long wavelength (Green)
  • POS B = Short wavelength (Blue)

Filter Wheel SPEC

  • POS A = Short wavelength (50-70um)
  • POS B = Long wavelength (70-100um)

Calibration sources

  • CS1
    • 48 Ohm = 55K
    • Chopper position center : - TBD
  • CS2
    • 58 Ohm = 60K
    • Chopper position center : + TBD

XY stage

  • Scale: 1" on the sky = 0.18mm
  • Spectrometer
    • Focus : +5mm (from FMILT-1, not incompatible with spectro FMILT-2)
    • Center spectro field :
      • x = -12.8 mm
      • y = +1.2 mm
  • Photometer
    • Focus : +5mm (from FMILT-1)
    • Center field :
      • x = 4 mm
      • y = 0

Herschel: FmIltProc/ReferenceNumbers (last edited 2009-07-15 14:32:37 by localhost)