FM ILT reference numbers
Ge:Ga detector settings
- Blue
par
value
comment
Bias d
198 mV
From bias / heater loop test
Bias r
10 mV
Component level value - to be verified on CRE checkout test
Heater
1.5 mA
Result from bias / heater loop test (2.68K/1.84mA) exceeds thermal budget
- Red
par
value
comment
Bias d
69 mV
From bias / heater loop test
Bias r
10 mV
Component level value - to be verified on CRE checkout test
Grating
- Configuration table: CONFGrat. Column : FMILTFast / FMILT_R_Fast for redundant
- Allowed working range :
- 45K - 1045K (nominal)
- 65K - 1065K (redundant)
- Conversion wavelength - grating position:
Chopper
- !!! Do NOT use the chopper in redundant mode !!!
- Configuration table CONFChopper / column : CSLopt3-600
- Accessible range : -23000 to +23000
- At +/- 25000 a field plate limit disables the controller
- Conversion chopper position - angle on the sky - FPU angle :
FPU Angle
Sky Angle
Chopper pos
Comment
- 8.5 deg
-6.32 arcmin
-23382
center CS1
-4 deg
-2.97 arcmin
-12101
0 deg
0 arcmin
0
Mechanical zero
+4 deg
2.97 arcmin
12431
+ 8.5 deg
6.32 arcmin
22347
center CS2
Other angles - in the PCSS (release >20070303A) :
from all import * # FPU angle : offset 650 print (650??? +) fpuAngle2ChopPosPol(Double1d([-4, 0, 4])) # Sky angle : scaling factor 80.69 print ((650??? +)fpuAngle2ChopPosPol(Double1d([-4, 0, 4])))/80.69 650???: optical zero is not known yet, must first be defined! you get angles referred to the mechanical zero by using fpuAngle2ChopPosPol or chopPos2FpuAnglePol
Filter Wheel PHOT
- POS A = Long wavelength (Green)
- POS B = Short wavelength (Blue)
Filter Wheel SPEC
- POS A = Short wavelength (50-70um)
- POS B = Long wavelength (70-100um)
Calibration sources
- CS1
- 48 Ohm = 55K
- Chopper position center : - TBD
- CS2
- 58 Ohm = 60K
- Chopper position center : + TBD
XY stage
- Scale: 1" on the sky = 0.18mm
- Spectrometer
- Focus : +5mm (from FMILT-1, not incompatible with spectro FMILT-2)
- Center spectro field :
- x = -12.8 mm
- y = +1.2 mm
- Photometer
- Focus : +5mm (from FMILT-1)
- Center field :
- x = 4 mm
- y = 0