1. Synchronization of Grating, Chopper & Readouts

The test just consists in operating the wavelength switching OBCP for a few minutes

Note : If possible, as soon as possible, replace steps 1 - 4 by SetupSpectroscopy_OBS

Script

TCL

Parameters

Comments

Duration [sec]

1

PACS_Spec_HK_Setup_OBS

Sets HK List to SPEC BOTH

2

CONF_grating_OBS

FMILTFast

3

CONF_chopper_OBS

TBD

TBVerified: Is this Script OK, which parameters should be used ?

4

PACS_GeGa_SFT_Init_Cold_He2_OBS

default

Swon & setup CREs

5

SPEC_CRE_setup_OBS

default

CRE Setup

6

SPEC_spu_setup_OBS

default

SPU Setup & Science Dataflow

7

OBCP_spec_freq_switch_OBS

P1=25 P2=1000 P3=1 P4=600000 P5=3 P6=600800 P7=599200 P8=P10=1 P9=-23382 P11=22347

Wavelength switching

374

8

SPEC_spu_reset_OBS

9

PACS_SAFE_MODE.tcl