1990
Comment:
|
1990
|
Deletions are marked like this. | Additions are marked like this. |
Line 14: | Line 14: |
|| 3 || CONF_phot_fltw_param_OBS || Yes || threshold || || | || 3 || CONF_phot_fltw_param_OBS || Yes || rate,drive_current,threshold || || |
Line 27: | Line 27: |
|| 3 || CONF_spec_fltw_param_OBS || Yes || threshold || || | || 3 || CONF_spec_fltw_param_OBS || Yes || rate,drive_current,threshold || || |
1. Filter Wheel control loop procedures
2. Conditions
All tests described can be executed from SAFE MODE of from SPECTROSCOPY.
3. 1. Treshold parameter optimisation Photometer filter wheel
A loop over steps 3 & 4 can be implemented for a number of thresholds
|
Script |
TCL |
Parameters |
Comments |
1 |
PACS_Phot_HK_Setup_OBS |
|
|
|
2 |
PACS_Phot_Fil_Diaghk_Setup_OBS |
|
|
|
3 |
CONF_phot_fltw_param_OBS |
Yes |
rate,drive_current,threshold |
|
4 |
PACS_Phot_Fil_nturns_OBS |
Yes |
nturns : number of full turns of the FW |
|
5 |
PACS_Diaghk_Reset_OBS |
|
|
|
6 |
PACS_SAFE_MODE.tcl |
|
|
|
4. 2. Treshold parameter optimisation Spectrometer filter wheel
A loop over steps 3 & 4 can be implemented for a number of thresholds
|
Script |
TCL |
Parameters |
Comments |
1 |
PACS_Spec_HK_Setup_OBS |
|
|
|
2 |
PACS_Spec_Fil_Diaghk_Setup_OBS |
|
|
|
3 |
CONF_spec_fltw_param_OBS |
Yes |
rate,drive_current,threshold |
|
4 |
PACS_Spec_Fil_nturns_OBS |
Yes |
nturns : number of full turns of the FW |
|
5 |
PACS_Diaghk_Reset_OBS |
|
|
|
6 |
PACS_SAFE_MODE.tcl |
|
|
|