= Preliminary analysis disconnecting redundant harness filter wheel = == Data == * Photometer filter wheel, redundant harness disconnected, treshold=2000 * 20070307/FILT_FWPHOT_test_disconn_20070307_01.tm * Photometer filter wheel, redundant harness disconnected, treshold=4000 * 20070307/FILT_FWPHOT_test_disconn_20070307_02.tm * Photometer filter wheel, short-circuit connector on redundant harness, treshold=2000 * 20070307/FILT_FWPHOT_test_disconn_short_20070307_01.tm * Photometer filter wheel, short-circuit connector on redundant harness, treshold=4000 * 20070307/FILT_FWPHOT_test_disconn_short_20070307_02.tm * Spectrometer filter wheel, redundant harness disconnected, treshold=2000 * 20070307/FILT_FWSPEC_test_disconn_20070307_01.tm * Spectrometer filter wheel, short-circuit connector on redundant harness, treshold=2000 * 20070307/FILT_FWSPEC_test_disconn_short_20070307_02.tm == Preliminary analysis by PR == * Disconnecting the redundant chain does not improve the situation at all (if we really had to work like that, I would recommand to try a threshold of 6000). * short circuit of the connectors brings us back to a "smooth" situation: smooth movement, and apparently much better reproducibility of stop positions. == Plots == * {{attachment:20070307_01_FWPHOT_Disconnect_HALLAB_time_move2_2000.png}} * {{attachment:20070307_01_FWPHOT_Disconnect_POSAB_time_all_2000.png}} * {{attachment:20070307_01_short_FWPHOT_DisconnectShortCircuit_HALLAB_time_move2_2000.png}} * {{attachment:20070307_01_short_FWPHOT_DisconnectShortCircuit_POSAB_time_all_2000.png}}